Golgappa.net | Golgappa.org | BagIndia.net | BodyIndia.Com | CabIndia.net | CarsBikes.net | CarsBikes.org | CashIndia.net | ConsumerIndia.net | CookingIndia.net | DataIndia.net | DealIndia.net | EmailIndia.net | FirstTablet.com | FirstTourist.com | ForsaleIndia.net | IndiaBody.Com | IndiaCab.net | IndiaCash.net | IndiaModel.net | KidForum.net | OfficeIndia.net | PaysIndia.com | RestaurantIndia.net | RestaurantsIndia.net | SaleForum.net | SellForum.net | SoldIndia.com | StarIndia.net | TomatoCab.com | TomatoCabs.com | TownIndia.com
Interested to Buy Any Domain ? << Click Here >> for more details...


why RID detector is used for selective materials like
Mannitol, Hydroxyethyl Starch and Sorbitol?

Answers were Sorted based on User's Feedback



why RID detector is used for selective materials like Mannitol, Hydroxyethyl Starch and Sorbitol?..

Answer / satheesh kumar

Since RI detectors useful for type of compounds such as
carbohydrates (Non-UV Absorbing Substance)Also,Each
compounds have it's own refractive Index.Hence it is useful
for Non-UV Absorbing substances.

Secondly,It's applicable only in isocratic mode.

Is This Answer Correct ?    7 Yes 0 No

why RID detector is used for selective materials like Mannitol, Hydroxyethyl Starch and Sorbitol?..

Answer / m.sivalingeswararao

These compounds are Saturated hydrocarbon products.
As UV-Light absorption is found in Unsaturated compounds not in saturated compounds,hence need to go for another appropriate technology, here it is RID

Is This Answer Correct ?    4 Yes 0 No

why RID detector is used for selective materials like Mannitol, Hydroxyethyl Starch and Sorbitol?..

Answer / jeff

because those chemicals don't have UV absorption or very low
absorption in UV range.

Is This Answer Correct ?    3 Yes 0 No

why RID detector is used for selective materials like Mannitol, Hydroxyethyl Starch and Sorbitol?..

Answer / sivanseyal g

detectors (RID) are widely used for many substances, such as to analyze components with low UV absorption, but it requires time for the baseline to stabilize,

Is This Answer Correct ?    0 Yes 0 No

why RID detector is used for selective materials like Mannitol, Hydroxyethyl Starch and Sorbitol?..

Answer / t

Hi

Is This Answer Correct ?    0 Yes 10 No

Post New Answer

More Analytical Chemistry Interview Questions

how to convert normal phase hplc to reverse phase and vice versa?

1 Answers  


why results are variation in same samples analysis in gc and hplc ?

2 Answers  


WHAT IS THE LIQUID PHASE IN DB-624 COLUMN IN GCHS

3 Answers  


Describe your experience with analytical instrumentation?

3 Answers   Alembic, Apotex, Aurobindo, DLF, Enal Drugs, Lupin, Natco, Pharma,


How require to perform linearity as per ANVISA? What are the acceptance criteria?

0 Answers  


difference between selectivity and specificity

2 Answers   Troikaa Pharmaceuticals,




0 Answers  


for which type of applications gas chromatography is used?

2 Answers  


In the method validation proceedure,lod & loq calculation we use the formulae for LOD=3.3*SD/slope and for LOQ=10*SD/slope,why we use 3.3 & 10 in the formulae?

1 Answers  


why we are using sodium,potassium phosphates for mobile phase preparation?

1 Answers  


Why potassium dichromate is used in calibration of uv?

0 Answers  


what is the difference between IR and FTIR instrument analysis

2 Answers  


Categories
  • Organic Chemistry Interview Questions Organic Chemistry (342)
  • Inorganic Chemistry Interview Questions Inorganic Chemistry (123)
  • Analytical Chemistry Interview Questions Analytical Chemistry (1387)
  • Physical Chemistry Interview Questions Physical Chemistry (62)
  • General Chemistry Interview Questions General Chemistry (478)
  • Chemistry AllOther Interview Questions Chemistry AllOther (190)