Why we use caffeine for HPLC calibration in Detector wave
ength accuracy test
Answer Posted / srinivas
caffeiene having different wavelents it having 2 maximas
and 1 minima.
1st maxima is 205nm
2nd maxima is 273nm
minima is 245nm
and it is primary reference standard and also suggested in
pharmacopiea.
It is very stable, we can use a long time foranalysis
i.e. it is not degradable
| Is This Answer Correct ? | 0 Yes | 0 No |
Post New Answer View All Answers
How to determine water content of bis tetrazole amine mono ammonium salt
How do we get end points and how many end points are possible for citric acid and di-acid not theorotically answer should be given practically.
Ratio of Water and Acetonitrile(diluent) is not recommended to use in cleaning method validation,why?
[3/28, 20:52] Manoj P Venkatpurwar: how many impurities require to inject in assay specificity that how we can find out? and in Rs also how?
how decide the clining method and cleaning method validation require for this perticular products?
How can the GC or HPLC method is selected to determine the impurity profile in drug product?
AT WHAT CONDITIONS WILL YOU ANALYSE A SAMPLE,WHICH ARE PRESERVED AT OTHER THAN THE AMBIENT CONDITIONS?(ie COLD STORAGE SAMPLES BELOW 20* CENTIGRADE?
Why use only methanol,used determinationof moisture content not use other solvents
Why we use potassium dichromate in uv calibration Exact reason behind it??
Why to use propyl paraben in HPLC gradient calibration
which are the diffrent batches in the pharmaceuticals?
Describe your field experience sample type collected,sample techniques,field measurements taken and equipment used?
how to calculate elemental impurities?
From where require to take the RLD sample?
What is dose dumping? why require to do?