Golgappa.net | Golgappa.org | BagIndia.net | BodyIndia.Com | CabIndia.net | CarsBikes.net | CarsBikes.org | CashIndia.net | ConsumerIndia.net | CookingIndia.net | DataIndia.net | DealIndia.net | EmailIndia.net | FirstTablet.com | FirstTourist.com | ForsaleIndia.net | IndiaBody.Com | IndiaCab.net | IndiaCash.net | IndiaModel.net | KidForum.net | OfficeIndia.net | PaysIndia.com | RestaurantIndia.net | RestaurantsIndia.net | SaleForum.net | SellForum.net | SoldIndia.com | StarIndia.net | TomatoCab.com | TomatoCabs.com | TownIndia.com
Interested to Buy Any Domain ? << Click Here >> for more details...

Why do we use 500ppb sucrose and 500ppb 1,4 benzoquinone in
TOC calibration?

Answer Posted / dev rawat

In Toc Analyzer Calibration (System suitability) we are preparing Standard sucrose of 500PPb and Benzoquinone Preparing of 500ppb . Sucrose is easy to oxidized and p- Benzoquinone is hard to decompose(oxidized).Its molecule bound not easy to oxidized so the response efficiency defined in USP (united state pharmacopoeia )its response efficiency not less then 85% and not more then 115% . BQ run as control sample in the References of Sucrose .And Its Calculation methods [(rss-rw ) / (rs-rw)]100.

Is This Answer Correct ?    15 Yes 0 No



Post New Answer       View All Answers


Please Help Members By Posting Answers For Below Questions

in which situation ion pair agent require to use?

799


when we talk about change in entropy we consider total entropy means of surround +entropy of system but when we talk about change in Gibss free energy we consider entropy of system.explain with reason.

2096


why xterra column require to use at higher ph?

968


what is partion and column chromatography

2468


How require to perform linearity as per ANVISA? What are the acceptance criteria?

1400


can i use hplc detector to uplc and why?

2347


if identification threshold crosses the limits then what next step?

915


what is the procedure for cleaning of lenses of hatr accesory of ftir instrument?

2338


why cone formation during dissolution?

847


In the HPLC Calibration done wavelength accuracy done between 200nm-280nm .but not done remaining 300-400nm not done ?

974


what is diffrence between specificity and selecivity?

837


calibrtion procedue of LC-MS

2760


before starting analytical method valodation what you checking? and how giving preference to start validation?

861


Why do we use KMnO4 in the test of control of obsorbance ? and why do we take specific quantity i.e 57-63mg?

2890


In which situation we require to prepare the standard solution from sample in Related substance method?

825