what split ratio in gas chromatography.
Answer Posted / khaja mohinuddin
Capillary columns are easily overloaded by injecting High Amount of Sample Concentration.
To Reduce or overcome this problem Split Ratio is Incorporated The Main use of this Split Ratio is Inject Less amount of Sample of interest into column or passes through it.
The Q? comes in a mind How it could be Done
Split vent is option Provide by Every manufacture like Agilent and shimadzu how it works i will expalin
Example :- ur method shows a split ratio of
1:30 Metan that 150ml/min of Split vent flow and Colum flow 5ml/min If u have Installed 0.53mm id Colum of any pahse.
Similarlly 1:50 mean 250ml/min of split ventflow and 5ml/min of column flow
formula split ratio = split vent flow / column flow
if know any of the parameter u can calculate by substituting in the above formula ...
thanks
| Is This Answer Correct ? | 7 Yes | 0 No |
Post New Answer View All Answers
how require to select dissolution media? what is discrimination?
in dissolution if one bowl got 70 percent 2nd bowl got 80 percent and 3rd bowl got 90 percent then how proceed?
how require to fix the linearity and accuracy cincentration range?
why cone formation during dissolution?
how can give the expiry period and restadardisation of volumetric solution
in which situation ion pair agent require to use?
What is similarity factor?
on the basis of bcs class how to proceed the solubility?
For limit test of heavy metals in BP, Method C require that the substance is ignited at a temperature not exceeding 800 °C. Why confines such the temperature?
Many times I don't got a caffeine peak in calibration of hplc using guard column ❓
Hi sir if any product monograph not given known impurities (may be 5 impurities) specifications then how we require to proceed for inhouse formulation?
what is lod and loq ?,why use k2cr2o7 , kcl h2so4 in uv calibration ?,why use benzophenone & caffene acetone in hplc calibration ?,what is leading peak in hplc ?why we do the calibration of limit of stry light in hplc & uv ?
give clarity of linearity and range in method validation
analytical method validation require to with respect to release specification or shelf life specification?
if tech transfer fails whay require to do?