How can you calculate manually "Slope" in pH calibration?
Answer Posted / chmreddy
A good electrode has a mV of 0 ± 30 mV in pH 7.
The slope should be 90 to 105%.
Example:
Readings
pH 4 : 4.29
pH 7 : 7.43
Convert pH 7 reading to mV
7 – 7.43 = -0.43 pH * 59.16 mV/pH = -25.44 (this value is okay because it is within 0 ± 30mV)
Convert pH 4 reading to mV
7 – 4.29 = 2.71 pH * 59.16 mV/pH = 160.32
Calculate slope
160.32 – (-25.44) = 185.76 mV/ 177.48 mV * 100 = 105 %
(this value is borderline as 105% is the upper limit)
| Is This Answer Correct ? | 13 Yes | 11 No |
Post New Answer View All Answers
In BP2013, Loperamid HCl monograph. Assay by titration with 0.1N sodium hydroxide using hydrocloric acid 0.01N and reading the volume added between the 2 points of inflexion. I have a question that if the diluent solvent is ethanol is certainly consumed a amount of volume of titrant, so this volume must be eliminated on the result calculation or not apart from first point which is subtracted above.
Why do we use KMnO4 in the test of control of obsorbance ? and why do we take specific quantity i.e 57-63mg?
How to do regeneration of Metacarb Pb plus column?
For standardization of volumetric solutions, acceptance limit for concentration is considered as 10% of nominal concentration and %RSD as NMT 0.2% (for appropriate no. of replicates). But , if %RSD has not met, what should be further action?
why require the ph, buffer during hplc mobile phase?
What is aggregate and fragments in SEC?
why require to add enzymes in capsules dissolution only? is it require to tablet also?
How would you decide dissolution medium for NCE compound of class I drug
i want model written test paper for the post of analyst in IOCL(indian oil corporation ltd) any one pls help me
how to develop the icp ms method? Application of icp ms?
If we do accuracy at same concentration at which linearity planned,what is the need to do linearity separately?
How to fix the sample conc. in Related substances test by HPLC?
using gradient pressure in gas chromatography are not ?using gradient pressure why
can we use the same detector in HPLC as well GC and what could be the differences we can find in the final chromato graph in any aspects?
Why sodium hydroxide used for maintain pH of phosphate buffer