How to calibrate a orifice differential pressure flow
transmitter in process?
Answer Posted / ayaz
first communicate ccr tag conform and turn MOS if required
when 5 way manifold
1 hp side close
2 equalizer open
3 lp side close
and then hp and lp drain or vent should be open
the transmitter will show zero if not calibrate it
first adjust the zero
2nd second span check and adjust
and then service
1 drain or vent close
2 equalizer close
3 and then hp side open and lp side open
| Is This Answer Correct ? | 3 Yes | 0 No |
Post New Answer View All Answers
I want to know the questions asked in ONGC interview(after being selected in the written exam) for Instrumentation branch.Kindly help.
Full form of KKS when use DCS.
Hi, Is there any smart plant instrumentation design training institute in Bangalore?
I want measure the volumetric flow of gas in waste heat recovery Boiler?Tell me how and which type of instruments required. Plz mail me:naba02222@gmail.com Sam please post the answer .
pls give me presant awerness quistion and answer in my emailid.
How flow can be measured with dp?
Regarding , there is an air cooler which control the fans(50% of AV FANS) by temperature in outlet , and the louver of Air cooler is controlled by HIC from control system , So if on Air cooler we will consider 2 individual I/P converter or positioner?
What is Span Error
Where are oxygen analysers installed in cfbc boilers? What are their usual readings?
For crude application in floating roof tank, how to calibrate a DP Transmitter? How is the floating roof weight considered?
How many ohms are required for proper earthing?
Which type of diaphragm use in costic line, dm water line, natural gas line, oil line?
our kks tagging is crossing more than 12 character please suggest how to take tags in the DCS
what deference between panel earth and systemearth
How to calibrate RC Feeder system in thermal power station?,If any instrument available to calibrate it,ple inform me. make:stock,