Golgappa.net | Golgappa.org | BagIndia.net | BodyIndia.Com | CabIndia.net | CarsBikes.net | CarsBikes.org | CashIndia.net | ConsumerIndia.net | CookingIndia.net | DataIndia.net | DealIndia.net | EmailIndia.net | FirstTablet.com | FirstTourist.com | ForsaleIndia.net | IndiaBody.Com | IndiaCab.net | IndiaCash.net | IndiaModel.net | KidForum.net | OfficeIndia.net | PaysIndia.com | RestaurantIndia.net | RestaurantsIndia.net | SaleForum.net | SellForum.net | SoldIndia.com | StarIndia.net | TomatoCab.com | TomatoCabs.com | TownIndia.com
Interested to Buy Any Domain ? << Click Here >> for more details...

what is difference of uv and pda

Answer Posted / rajib hazarika

1. By PDA detector, we can measure the area/height
of particular sample at wavelengths ranging from
190 to 800 nm by a single injection( 3D scan is possible(Response Vs time Vs wavelength). Whereas in uv detector scanning is not possible.)
2. for unknown sample with unknown lambda maximum with a single injection PDA we can determine the lambda maximum but by UV several injection is required with different wavelength.
(for method validation, process validation, Research and development PDA is prefered as compared to UV
3. If proper elution is established with the help of multi step gradient mobile phase and column then analysis of mixture of compounds(with different Lambda max) can be possible by a single injection in PDA.

Is This Answer Correct ?    8 Yes 1 No



Post New Answer       View All Answers


Please Help Members By Posting Answers For Below Questions

USP methodology, EP methodology, IP methodology, among three if possible to use one methodology for qualify working standard to use USP, EP, IP ? Please explain...

2073


Please give idea about method development for cleaning method and how maco level establish

797


UV and PDA detector, which have less signal to noise ratio?

1177


What is Band broading effect explain

1941


in gas chromatography what is the difference between gas flow rate and average linear velocity ?

2654


on the basis of bcs class how to proceed the solubility?

833


what is procedure of actual procedure for registration of drug?

2055


what is mean by covalidation

833


What is the accrptance criteria in RSD for RS method precision on basis of impurity percentages?

887


what you do when compound having dwel wavelength?

808


please explian me about area normalisation method,diluted standard method,impurity vs impurity and which one has to be selected in method development

1691


In HPLC Calibration, On which basis RSD Limit of noise test is fixed (NMT 33.0 % )

3552


what is the extinction coefficient for Indomethacin or at 319nm wave length.

4127


How they found 1mL of K.F reagent is equivalent to 5mg of water and if we change the composition of K.F reagent, is it can neutralize more amount of water?

1830


how to select short coloum and long coloumn for new molecule

2527