what is difference of uv and pda

Answer Posted / rajib hazarika

1. By PDA detector, we can measure the area/height
of particular sample at wavelengths ranging from
190 to 800 nm by a single injection( 3D scan is possible(Response Vs time Vs wavelength). Whereas in uv detector scanning is not possible.)
2. for unknown sample with unknown lambda maximum with a single injection PDA we can determine the lambda maximum but by UV several injection is required with different wavelength.
(for method validation, process validation, Research and development PDA is prefered as compared to UV
3. If proper elution is established with the help of multi step gradient mobile phase and column then analysis of mixture of compounds(with different Lambda max) can be possible by a single injection in PDA.

Is This Answer Correct ?    8 Yes 1 No



Post New Answer       View All Answers


Please Help Members By Posting Answers For Below Questions

How can I develope method of dissolution by HPLC OR UV

1893


Why we check moisture,ash & AIA in product

1402


why we are using hexane in calibration of number of drop per mL

1170


USP methodology, EP methodology, IP methodology, among three if possible to use one methodology for qualify working standard to use USP, EP, IP ? Please explain...

1558


For standardization of volumetric solutions, acceptance limit for concentration is considered as 10% of nominal concentration and %RSD as NMT 0.2% (for appropriate no. of replicates). But , if %RSD has not met, what should be further action?

1249






If we do accuracy at same concentration at which linearity planned,what is the need to do linearity separately?

1755


My question about gas chromatography sulfur chemiluminsecence detector. I test unknown sample gas by GC-SCD (calibrated ) and the result of *H2S is 279 PPM , *but when I test the same sample with the GC-TCD (calibrated ) the value of *H2S is 2500 PPM . I'd like to inform you that both GCs are calibrated and have very good operation conditions with stable parameters . the question is if the sample gas with higher H2S over detection limits of SCD detector (1000 ppm). why the result it 279 ppm Best regards

1112


WHAT IS THE PURPOSE WE USE HCL, CHOPPER, BURNER, MONOCHROMATOR in AAS? PLEASE EXPLAIN BRIEFLY

2101


identification is for unknown? qualification for known? reporting for LOQ?

286


How to determine water content of bis tetrazole amine mono ammonium salt

2966


4. Describe the operation of the Craig apparatus. Chapter.3 Equilibrium Processes in Separations

2157


why holium oxide prepared in 10% HCLO4 SOLUTION

993


Which are the diffrent grades of api in pharma?

314


What is the principle of Thermal conductivity detector and FID?

2259


What is the difference between purge septum flow and column flow in gas chromatography?

1837