Why we more noise in PDA detector compared to UV detector?

Answer Posted / anil

There is no Integration function in the conventional UV-Vis spectrophotometer which
accumulates the signal. For example, the total required time will be 1000 sec. for 1000 data
points and it takes 1 sec. to measure one datum. In this case, all 1000 data have the same signal
to noise ratio (S/N).
But in a PDA instrument which has a 1000 photodiode array, 1000 data points can be
measured in 1 sec. and it would take 1/1000 sec. to achieve the same result obtainable in 1 sec.
in a conventional instrument. Therefore, when the same sample is measured for 1000 sec in a
PDA instrument, the signal is accumulated and is 1000 times greater than when measuring for 1
sec. The noise will be 10001/2
. This means that the S/N ratio is improved by 10001/2
.
This resulting benefit of fast data acquisition is termed Felgett’s S/N Advantage or
Multichannel Advantage

Is This Answer Correct ?    10 Yes 1 No



Post New Answer       View All Answers


Please Help Members By Posting Answers For Below Questions

Explain the relations between number of carbon atoms in alkanes and retention time ?

1630


WHAT IS THE USE OF STARTER ?????????

1913


How do we quantify crystaline and amarpous forms by using (NMR, XRD)spectroscopic techniques? Which any others instruments are useful for this quantification? explain

1958


how require to set assay concentration for standard and sample?

373


How will u perform degradation study in detectors other thann PDA in HPLC Methodology

2022






What is the main difference between ODS1 and ODS2 Hplc column.

4477


What is similarity factor?

731


In the HPLC Calibration done wavelength accuracy done between 200nm-280nm .but not done remaining 300-400nm not done ?

463


What is similarity factor and its use in api strength?

801


if you given one product then which tests you will perform?

323


why salisylic acid not using now days for disdolution calibration?

305


What are usp limits for theoritical plats,resoution,tailing factor,peak to valley ratio

1343


Why perchloric acid used for potentiometric titration

2169


WHAT IS THE CRITERIA FOR SELECTION OF TIME OF DISSOLUTION AND THE MEDIUM OF DISSOLUTION?

2145


How to calculation of residual solvent methanol (3000 ppm) standard solution, and how many quantity used of test sample

2543