Golgappa.net | Golgappa.org | BagIndia.net | BodyIndia.Com | CabIndia.net | CarsBikes.net | CarsBikes.org | CashIndia.net | ConsumerIndia.net | CookingIndia.net | DataIndia.net | DealIndia.net | EmailIndia.net | FirstTablet.com | FirstTourist.com | ForsaleIndia.net | IndiaBody.Com | IndiaCab.net | IndiaCash.net | IndiaModel.net | KidForum.net | OfficeIndia.net | PaysIndia.com | RestaurantIndia.net | RestaurantsIndia.net | SaleForum.net | SellForum.net | SoldIndia.com | StarIndia.net | TomatoCab.com | TomatoCabs.com | TownIndia.com
Interested to Buy Any Domain ? << Click Here >> for more details...

Why we more noise in PDA detector compared to UV detector?

Answer Posted / anil

There is no Integration function in the conventional UV-Vis spectrophotometer which
accumulates the signal. For example, the total required time will be 1000 sec. for 1000 data
points and it takes 1 sec. to measure one datum. In this case, all 1000 data have the same signal
to noise ratio (S/N).
But in a PDA instrument which has a 1000 photodiode array, 1000 data points can be
measured in 1 sec. and it would take 1/1000 sec. to achieve the same result obtainable in 1 sec.
in a conventional instrument. Therefore, when the same sample is measured for 1000 sec in a
PDA instrument, the signal is accumulated and is 1000 times greater than when measuring for 1
sec. The noise will be 10001/2
. This means that the S/N ratio is improved by 10001/2
.
This resulting benefit of fast data acquisition is termed Felgett’s S/N Advantage or
Multichannel Advantage

Is This Answer Correct ?    10 Yes 1 No



Post New Answer       View All Answers


Please Help Members By Posting Answers For Below Questions

What is the formula to determine the concentration of M of a solution given the % transmittance? Use %T = 43.7 as an example.

2753


what is biorelivent dissolution media?

856


what is the difference between potentiometric titration and karl fischer titration?

3257


what is procedure of actual procedure for registration of drug?

2055


which are the diffrent batches in the pharmaceuticals?

861


WHICH STANDARD WE USE IN SILICA ANALYSIS IN SILICA-GEL METHOD FOR SOLID SAMPLE

2183


how you start the solubility study?

861


What is shaking level in GC?

833


WHAT IS THE CRITERIA FOR SELECTION OF TIME OF DISSOLUTION AND THE MEDIUM OF DISSOLUTION?

2624


if your impurity coeluting with each other in that situation how require to set specification? is it acceptable?

1042


why we are using hexane in calibration of number of drop per mL

1790


How to decide assay range for non pharmacopeial API analysis by HPLC? Could you give me any reference for same e.g Guidelines or Paper publications?

3397


Why to use propyl paraben in HPLC gradient calibration

7580


Identify problem faced when mass spectroscopy is used with HPLC system ?

1925


what are the standard methods for HPLC?

2829