WHAT IS PVT TEST IN DISSOLUTION TEST , WHAT IS PVT SIGNIFICANCE AND HOW IT IS PERFORMED ?

Answer Posted / archana

PVT Test in dissolution test means performance verification test which is also known as apparatus suitability test

Frequency of this test is once in six months.

This can be done by using prednisone USP Tablet 10 mg /tablet
for disintegrating type of tablets.

Current lot for this tablet is O0C056.

For Type 1 Apparatus
Dissolution medium :900 ml water
Temp :37°C
RPM :50
%dissolved :53-77%(O0C056)
Time :30mins
For Type 2 Apparatus
Dissolution medium :900ml water
Temp :37°C
RPM :50
%dissolved :27-48%(O0C056)
Time :30mins
Readings r taken at 242 nm
Std concentration 20 mg /ml of prednisone USP

Salicylic Acid Tablets are used for non disintegrating type of tablets
Current lot is Q0D200

For Type 1 Apparatus
Dissolution medium :900 ml of phosphate buffer of pH 7.4+/-.005
Temp :37°C
RPM :50
%dissolved :23-29%(Q0D200)
Time :30mins
For Type 1 Apparatus
Dissolution medium :900 ml of phosphate buffer of pH 7.4+/-.005
Temp :37°C
RPM :50
%dissolved :17-26%(Q0D200)
Time :30mins
Readings r taken at 296 nm
Std concentration 75 mg /ml of Salicylic acid USP

Is This Answer Correct ?    2 Yes 0 No



Post New Answer       View All Answers


Please Help Members By Posting Answers For Below Questions

UV and PDA detector, which have less signal to noise ratio?

645


Hi sir if any product monograph not given known impurities (may be 5 impurities) specifications then how we require to proceed for inhouse formulation?

280


I have compare C2H2-air and C2H2-N2O flame AAS on determination calcium. Both use same range of std to plot calibration curve. (2-6ppm) When i measure the sample with phosphate, KCl and LaCl, C2H2-N2O flame give false positive result, around 0.5ppm. When i measure the sample with phosphste, KCl and EDTA. C2H2-N2O flame also give 0.5ppm false positive. But both above mentioned sample would not give false positive when measured by C2H2-air flame. What is the reason?

2976


In HPLC calibration, caffeine is used as primary standard for wave length calibration due to caffeine is having dual maxima at 273 & 205 nm and one minima at 245 nm. Any body can give reference of these details from any pharmacopeia (with chapter no.) or any other authentic guideline?

1453


If vendor having more imp than monograph then how to proceed? and how to set spec?

336






function of detecter in hplc ,gc and spectroscopy? function of carrier gas in gc?

2066


how decide the clining method and cleaning method validation require for this perticular products?

299


What is the main difference between ODS1 and ODS2 Hplc column.

4475


In GC calibration,what is the procedure to calibrate the TCD (thermal counductivity detector)suggest ?

2709


Why only Copper standard is used to calibrate Atomic Absorption spectrophotometer?

2137


How to set analytical specification for combination products?

324


As per ICH related substances stability trend limit from initial to shelf life

2665


We use Potassium Dichromate solution for the calibration of UV-Visible Spectrophotometer in UV region. My question is to calibrate visible region which solution could be used in photometric accuracy

1317


sop of a uv visible spectrophotometer double beam elico model

5353


how require to set assay concentration for standard and sample?

364