why do we adjusted ph in the mobilphase
Answer Posted / simhadri
It is better to adjust our mobile PH minium 1 or 2 units away from the analyte PKa,to get good peak shape and good retention.
| Is This Answer Correct ? | 8 Yes | 1 No |
Post New Answer View All Answers
why require the ph, buffer during hplc mobile phase?
what is partion and column chromatography
effect of pore size, pore volume, partical size, column length, carbon load on retention time? what is carbon load? what is the use?
can i use hplc detector to uplc and why?
How to calibration of the uv spectroscopy and its test?
What is the difference between chromatographic purity and related substances?
UV and PDA detector, which have less signal to noise ratio?
Colorimtry
why cone formation during dissolution?
I have compare C2H2-air and C2H2-N2O flame AAS on determination calcium. Both use same range of std to plot calibration curve. (2-6ppm) When i measure the sample with phosphate, KCl and LaCl, C2H2-N2O flame give false positive result, around 0.5ppm. When i measure the sample with phosphste, KCl and EDTA. C2H2-N2O flame also give 0.5ppm false positive. But both above mentioned sample would not give false positive when measured by C2H2-air flame. What is the reason?