Golgappa.net | Golgappa.org | BagIndia.net | BodyIndia.Com | CabIndia.net | CarsBikes.net | CarsBikes.org | CashIndia.net | ConsumerIndia.net | CookingIndia.net | DataIndia.net | DealIndia.net | EmailIndia.net | FirstTablet.com | FirstTourist.com | ForsaleIndia.net | IndiaBody.Com | IndiaCab.net | IndiaCash.net | IndiaModel.net | KidForum.net | OfficeIndia.net | PaysIndia.com | RestaurantIndia.net | RestaurantsIndia.net | SaleForum.net | SellForum.net | SoldIndia.com | StarIndia.net | TomatoCab.com | TomatoCabs.com | TownIndia.com
Interested to Buy Any Domain ? << Click Here >> for more details...

Explain about Low pressure gradient and High Pressure
gradient systems in HPLC?

Answer Posted / venkata narayana

Answer:
In a high pressure gradient pump, each different mobile phase is delivered by an individual pump head and then the mobile phases are mixed at the pump outlet. In a low pressure gradient pump, different mobile phases are mixed using a valve before entering the pump head.


As a result of the fact that the low-pressure gradient design uses only one pump head, it is of lower cost. It can also use more types of mobile phase without significant increase of cost.


Since solvent mixing point is much closer to the column head in the high-pressure gradient design, it provides a much faster gradient. This is measured using delay volume. The value can be 50-300 uL for high pressure gradient pump and can be 2 to 3 times larger for a low pressure gradient pump. A small delay volume is important when the analysis time is short or the flow rate is low. If the delay volume is too large, it become impossible to obtain reproducible gradient run since the planed composition cannot reach the column head before a run is finished.

Is This Answer Correct ?    20 Yes 1 No



Post New Answer       View All Answers


Please Help Members By Posting Answers For Below Questions

Why to use propyl paraben in HPLC gradient calibration

7449


What is the diference between residual solvents and organic volatile matter

2023


What is the difference between chromatographic purity and related substances?

1360


How require to perform linearity as per ANVISA? What are the acceptance criteria?

1325


what is the principle of UV Vis spectroscopy, AAS, ICP OES,ICPAES, ICP-MS and FTIR

9241


What is gelatinization?

1612


How can the GC or HPLC method is selected to determine the impurity profile in drug product?

7586


Which is the highly polar and highly non polar column in HPLC?

2297


Why are use silicon oil mr?

1194


we can use expired sample for validation and analyst qualification?

799


What is the acceptance criteria for moisture balance when calibrated with sodium sulfate

2034


How to calculation of residual solvent methanol (3000 ppm) standard solution, and how many quantity used of test sample

3226


WHAT IS THE PURPOSE WE USE HCL, CHOPPER, BURNER, MONOCHROMATOR in AAS? PLEASE EXPLAIN BRIEFLY

2603


WHY given much noise PDA dector then VWD Dector

2084


how the compound separate in coulunm,explain

2523