how calibrate the hplc detector

Answer Posted / m.sivalingeswara rao

Hplc detector calibration to be done to know the acuracy of
the emitted light wave length to the set value as well as
to check the sensitivity of the detecotr to respond with
the concentration of analyte.
To check accuracy generally caffeine solution is used
because it shows maxima and minima at particular wave
lengths(i.e maxima at 206nm,273nm and Minima at 241nm,in
this range we can check accuracy of lamp emitted light).
Sensitivity by permorming linearity test with different
concetration of caffeine solution.
Reproducibility by injecting the replicate injection of
Caffeine solution and measuring the %RSD of the responses

Is This Answer Correct ?    19 Yes 0 No



Post New Answer       View All Answers


Please Help Members By Posting Answers For Below Questions

In which situation we require to analytical method validation of excipient?

288


for heavy metal test lead used in which form pure form or any other form for preparation of lead standard sol

5292


How can we calculate "confidence interval" in analytical method validation? Pl. explain with example.

3595


In Dissolution Test why limit is define Q+5% what is the role of +5%.

2567


why we are using hexane in calibration of number of drop per mL

1165






What is viod volume and peak purity in HPLC?

2332


what is turbidimetric titration?Give 2 examples.

1985


What is split ratio in Gc? Splitless? how requirr to select?

323


Why acetonitrile and water are used as extraction solvent when analysis melamine? I thought they are miscible and won't be able to separate...

1216


What is third generation HPLC Columns?

1507


what is partion and column chromatography

1815


I have compare C2H2-air and C2H2-N2O flame AAS on determination calcium. Both use same range of std to plot calibration curve. (2-6ppm) When i measure the sample with phosphate, KCl and LaCl, C2H2-N2O flame give false positive result, around 0.5ppm. When i measure the sample with phosphste, KCl and EDTA. C2H2-N2O flame also give 0.5ppm false positive. But both above mentioned sample would not give false positive when measured by C2H2-air flame. What is the reason?

2973


Why Ethanol is Used for Standardization of GC Head space?

3716


if you have given a blank water how you analyse by general analytical techniques?

1504


what is the importance of peak purity in HPLC and how we can calculate through manul(not software)?

3985