Why we use caffeine for HPLC calibration in Detector wave
ength accuracy test
Answer Posted / kishanpal singh, fare labs pvt
1. caffeine easily available in market.
2. In general we all use caffeine as a standard for
calibration of HPLC specially for (Calibration of Wave
Length Accuracy) since it will scan the whole of the UV
range i.e It gives a multiwavelength response like
first maxima at 205 nm
second maxima at 273 nm &
minima at 245nm
3. It is very stable, we can use a long time for analysis
i.e. it is not degradable
4. In this case we can use Erbium Perchlorate which will
scan the whole range(200 nm to 600 nm).It is stable and
little costly but not easily available in market.
KISHANPAL SINGH
Is This Answer Correct ? | 1 Yes | 0 No |
Post New Answer View All Answers
What is the formula to determine the concentration of M of a solution given the % transmittance? Use %T = 43.7 as an example.
in each media we require to use SLS? how to proceed?
How we performed the force degradation for drug substance, is any specific guideline is available for each parameter(Acidic, basic, oxidation,heat)? what conditions you mentained for above parameters.
In HPLC (Waters) software Empower1,2 &3 what is the updation for each version
in which situation ion pair require to use?
How to set analyticl specification for combination products?
inhouse product is in capsule form in combination and RLD is in tablet form then can we proceed for multimedia CDP? in inhouse capsule product disso is paddle with sinker in release media is there then RLD product in tablet form then with same as paddle with sinker we can proceed n.a.?
why glutent are detected in the rice cereal baby food product even manufacturer claimed that they are using rice and milk only?we have using ELISA to do the test,and rice supposed not containing any glutent,rite?We already repeat the test so many times and it still detected.just wondering where the glutent came from?
How they found 1mL of K.F reagent is equivalent to 5mg of water and if we change the composition of K.F reagent, is it can neutralize more amount of water?
For titration in anhydrous media with perchloric acide, if lack of titrator, Which indicator is been used for replacement. How calculate pH of test solution to choose suitable indicator?
As per ICH related substances stability trend limit from initial to shelf life
what are the guidelines for analytical method validations?
EXPLAIN THE CALCULATION PART OF AN KF TITRATOR CALIBRATION WITH SUITABLE EXAMPLE
effect of pore size, pore volume, partical size, column length, carbon load on retention time? what is carbon load? what is the use?