Why we use caffeine for HPLC calibration in Detector wave
ength accuracy test
Answer Posted / kishanpal singh, fare labs pvt
1. caffeine easily available in market.
2. In general we all use caffeine as a standard for
calibration of HPLC specially for (Calibration of Wave
Length Accuracy) since it will scan the whole of the UV
range i.e It gives a multiwavelength response like
first maxima at 205 nm
second maxima at 273 nm &
minima at 245nm
3. It is very stable, we can use a long time for analysis
i.e. it is not degradable
4. In this case we can use Erbium Perchlorate which will
scan the whole range(200 nm to 600 nm).It is stable and
little costly but not easily available in market.
KISHANPAL SINGH
| Is This Answer Correct ? | 1 Yes | 0 No |
Post New Answer View All Answers
why digestion require in icpms?
iam usig ph buffers merk. manually how to prepare ?
How to set analytical specification for combination products?
in which situation require to use paddle and basket?
what is mean by dissolution biowaiver study?
We use Potassium Dichromate solution for the calibration of UV-Visible Spectrophotometer in UV region. My question is to calibrate visible region which solution could be used in photometric accuracy
why sre you used Potassium hydrogen phthalate in standarisation of 1N NaOH and 0.1 N Perchloric Acid?
what are risk assement in the analytical qbd?
Why sometime potassium ortho phosphate mix with acetone use for sample preparation for hplc
1)What's the meaning of Absorption,give a example. 2)What's the meaning of Adsorption,give a example. 2)what is the difference between Absorption and Adsorption.
What is the accrptance criteria in RSD for RS method precision on basis of impurity percentages?
My question about gas chromatography sulfur chemiluminsecence detector. I test unknown sample gas by GC-SCD (calibrated ) and the result of *H2S is 279 PPM , *but when I test the same sample with the GC-TCD (calibrated ) the value of *H2S is 2500 PPM . I'd like to inform you that both GCs are calibrated and have very good operation conditions with stable parameters . the question is if the sample gas with higher H2S over detection limits of SCD detector (1000 ppm). why the result it 279 ppm Best regards
Why perchloric acid used for potentiometric titration
what is the purge flow & how to calculate
In the HPLC Calibration done wavelength accuracy done between 200nm-280nm .but not done remaining 300-400nm not done ?