Golgappa.net | Golgappa.org | BagIndia.net | BodyIndia.Com | CabIndia.net | CarsBikes.net | CarsBikes.org | CashIndia.net | ConsumerIndia.net | CookingIndia.net | DataIndia.net | DealIndia.net | EmailIndia.net | FirstTablet.com | FirstTourist.com | ForsaleIndia.net | IndiaBody.Com | IndiaCab.net | IndiaCash.net | IndiaModel.net | KidForum.net | OfficeIndia.net | PaysIndia.com | RestaurantIndia.net | RestaurantsIndia.net | SaleForum.net | SellForum.net | SoldIndia.com | StarIndia.net | TomatoCab.com | TomatoCabs.com | TownIndia.com
Interested to Buy Any Domain ? << Click Here >> for more details...

How we can analysed unknown sample by hplc....

Answer Posted / balaji

first check the structure of the compound.identify the
chromophore groups(for UV Detector),refractive index status
(for RI Detector),reactive status with other chromophore
group compounds(Derivitation method for UV.then check the
solubility status of the compound.prepare a known
concentration solution.

In HPLC select a C18 Column which can serve better for the
polar and not polar compounds.select a buffer pH Oof +/-1
unit of pka value of the compound prepare a buffer mix with
some 20% of organic solvent to chnge the elution early
(based on polarity)

inject the prepared known concentration sample solution in
to above system by observing the chromatogram procced
further by altering buffer
pH,COLUMN,FLOW,TEMPERATURE,ORGANIC RATIO Sample size etc.

but to quantify the unknown concentration we need a known
concentration of standard.

Is This Answer Correct ?    2 Yes 1 No



Post New Answer       View All Answers


Please Help Members By Posting Answers For Below Questions

how pda detector works over uv?

737


how we can identify the impurity is coming below loq at transfering site?

863


[3/30, 13:29] Manoj P Venkatpurwar: How hplc column selection according to structure? How mobile phase buffer selection on molecule structure?

776


If we do accuracy at same concentration at which linearity planned,what is the need to do linearity separately?

2295


What should be the level/ Percentage of Arsenic in Arsenic free zinc.

2224


why salisylic acid not using now days for disdolution calibration?

764


What is diffrence between extractable volume and deliverable volume? Answer pls

4556


what is mean by covalidation

771


How to set analytical specification for combination products?

815


How to fix the sample conc. in Related substances test by HPLC?

1085


what is difference between UV - VISIBLE MODEL NO like 1600,1601,1700 etc ? plz explain me

2666


In Assy by potentio metry titration method how select MET U ,DET U, SET U,

2155


If change in specification which parameter require to do for validation? if change in chromtographic condition then which parameter? if api change then which parameter? if change composition then which parameter? if old method not work out then whicj parameter? if additional one impurity added then which parameter of validation require to do on above each conditions? elaborate separately

848


Colorimtry

1647


what is mean by ambient temperature?

912