Answer Posted / natraj kumar
Example HPLC calibration : - Suppose one parameter is
failed .keep the status board for Instrument breakdown.And
raise the OOC (Out of calibration )document take the
approval from QC and QA Head. After rectification or change
the part of the Instrument, check the failed parameter of
calibration.
If it is Passed, then check randomly impact assesment of
the previous analysis during this calibration period.If
there is no impact then u can proceed for regular
usage.Total process to be complete in 30 days only.
OOC document keep the calibration file. Service report is
also keep the filing internal or External service person.
Dr.Reddys Lab
Is This Answer Correct ? | 72 Yes | 3 No |
Post New Answer View All Answers
calibrtion procedue of LC-MS
for inorganic molecules require to do RS, Assay and disso?
Why Ethanol is Used for Standardization of GC Head space?
How can I develope method of dissolution by HPLC OR UV
how much mass should be there in volumetric flask while in preparation of sample for assay?
[3/28, 20:52] Manoj P Venkatpurwar: how many impurities require to inject in assay specificity that how we can find out? and in Rs also how?
In Dissolution Test why limit is define Q+5% what is the role of +5%.
How do we get end points and how many end points are possible for citric acid and di-acid not theorotically answer should be given practically.
Why dissolution test is not performed in all of the products
Identify problem faced when mass spectroscopy is used with HPLC system ?
What is delay volume?
If combination product how require to identify which imp is of which api?
in OSD forms require to use gas chromatography?
could negative ions be produced by bombardment process in mass spectrometry?
what is turbidimetric titration?Give 2 examples.