Golgappa.net | Golgappa.org | BagIndia.net | BodyIndia.Com | CabIndia.net | CarsBikes.net | CarsBikes.org | CashIndia.net | ConsumerIndia.net | CookingIndia.net | DataIndia.net | DealIndia.net | EmailIndia.net | FirstTablet.com | FirstTourist.com | ForsaleIndia.net | IndiaBody.Com | IndiaCab.net | IndiaCash.net | IndiaModel.net | KidForum.net | OfficeIndia.net | PaysIndia.com | RestaurantIndia.net | RestaurantsIndia.net | SaleForum.net | SellForum.net | SoldIndia.com | StarIndia.net | TomatoCab.com | TomatoCabs.com | TownIndia.com
Interested to Buy Any Domain ? << Click Here >> for more details...

In GC, What does 'Column condensation test' mean by?

Answer Posted / sri krishna

Condensation Test
Use this test whenever injector or carrier gas contamination
problems are suspected (e.g., ghost peaks or erratic baselines).

* Leave the GC at 40-50&#65533;C for 8 or more hours.
* Run a blank analysis (i.e., start the GC, but with no
injection) using the normal temperature conditions and
instrument settings.
* Collect the chromatogram for this blank run.
* Immediately repeat the blank run as soon as the first
one is completed. Do not allow more than 5 minutes to elapse
before starting the second blank run.
* Collect the chromatogram for the second blank run and
compare it to the first chromatogram.
* If the FIRST chromatogram contains a substantially
larger amount of peaks and baseline instability, then that
is an indication that there is contamination upstream of the
capillary column (ie. contaminated inlet, dirty carrier gas,
etc.).
* If BOTH chromatograms contain few peaks or very little
baseline drift, it can be assumed that the carrier gas
and/or inlet are relatively clean.
* If BOTH chromatograms contain a significant amount of
noise and/or baseline drift, then that usually is an
indication that the detector or detector gases are contaminated.

Is This Answer Correct ?    8 Yes 1 No



Post New Answer       View All Answers


Please Help Members By Posting Answers For Below Questions

cefoperozone and sulbactam inj. hplc test method

2691


which situation gc hs and gc als require to use?

787


if tech transfer fails whay require to do?

844


In HPLC calibration, caffeine is used as primary standard for wave length calibration due to caffeine is having dual maxima at 273 & 205 nm and one minima at 245 nm. Any body can give reference of these details from any pharmacopeia (with chapter no.) or any other authentic guideline?

1973


What is the formula to determine the concentration of M of a solution given the % transmittance? Use %T = 43.7 as an example.

2684


WHAT IS THE USE OF STARTER ?????????

2425


In Assy by potentio metry titration method how select MET U ,DET U, SET U,

2171


what is impurity profile. how to interpret this impurity profile to a drug product or drug substance.

2629


why should we perform dissolution PVT calibration only by UV spectrometer, not by HPLC ?

1708


why octanol used to determine the partition coefficient ?

2458


Home heating oil or furnace oil what is made from? Can (base oil plus diesel/kerosen) = home heating oil?

2184


Difference between the quantitative analysis and qualitative analysis?

1178


What is importance of pH in the Heavy metal test and How it effect on the test results?

2283


How can we confirm the HPLC column is end-capped or not? Is it possible to identify by physical appearance?

3218


how to calculate elemental impurities?

1032