injection volume accruacy test i calibration of hplc
Answer Posted / pratima bhandarkar
Connect union in to the HPLC.
Mobile phase: Methanol: Water (50:50)
Temperature : 25°C
Retention time :1min
Flow rate :0.5mL/min
Take one HPLC vial. Fill it with water. Weigh it (W1)
Make 6 injections with 50µl injection volume
Again weigh it (W2)
Mean volume = (W1-W2)/6*100 = µl
Limit = +-2% of the injection volume
| Is This Answer Correct ? | 8 Yes | 3 No |
Post New Answer View All Answers
Why use only methanol,used determinationof moisture content not use other solvents
what parameters we will consider while developing a HPLC method and how we confirm our HPLC method is valid?
Which are the diffrent grades of api in pharma?
What is the Formula for coreletion coefficient in plhplc calibration
EXPLAIN THE CALCULATION PART OF AN KF TITRATOR CALIBRATION WITH SUITABLE EXAMPLE
what you do when compound having dwel wavelength?
For standardization of volumetric solutions, acceptance limit for concentration is considered as 10% of nominal concentration and %RSD as NMT 0.2% (for appropriate no. of replicates). But , if %RSD has not met, what should be further action?
can we use the same detector in HPLC as well GC and what could be the differences we can find in the final chromato graph in any aspects?
why we use glass fiber filters use in some situation?
when we talk about change in entropy we consider total entropy means of surround +entropy of system but when we talk about change in Gibss free energy we consider entropy of system.explain with reason.
What should be the level/ Percentage of Arsenic in Arsenic free zinc.
how to develop the icp ms method? Application of icp ms?
why require to add enzymes in capsules dissolution only? is it require to tablet also?
what is the procedure for cleaning of lenses of hatr accesory of ftir instrument?
What is similarity factor and its use in api strength?