why a ph meter require KCL solution for their storing
condition?
Answer Posted / dr. a. bera, cdl
When not in use, the glass probe tip must be kept wet at all times to avoid the pH sensing membrane dehydration and the subsequent dysfunction of the electrode.
A glass electrode alone (i.e., without combined reference electrode) is typically stored immersed in an acidic solution of around pH 3.0. In an emergency, acidified tap water can be used, but distilled or deionised water must never be used for longer-term probe storage as the relatively ionless water "sucks" ions out of the probe membrane through diffusion, which degrades it.
Combined electrodes (glass membrane + reference electrode) are better stored immersed in the bridge electrolyte (often KCl 3 M) to avoid the diffusion of the electrolyte (KCl) out of the liquid junction.
Is This Answer Correct ? | 72 Yes | 5 No |
Post New Answer View All Answers
How can we calculate "confidence interval" in analytical method validation? Pl. explain with example.
How much is the standard area for glc analysis
Tell me something about Vitamin A test method by HPLC
i want model written test paper for the post of analyst in IOCL(indian oil corporation ltd) any one pls help me
about method developement in hplc
Please give idea about method development for cleaning method and how maco level establish
What is aggregate and fragments in SEC?
[3/30, 13:29] Manoj P Venkatpurwar: How hplc column selection according to structure? How mobile phase buffer selection on molecule structure?
In which situation we require to analytical method validation of excipient?
sop of a uv visible spectrophotometer double beam elico model
Identify problem faced when mass spectroscopy is used with HPLC system ?
in api coa contains only process impurities and in product coa contains degradation impurities?
analytical method validation require to with respect to release specification or shelf life specification?
[3/28, 20:52] Manoj P Venkatpurwar: how many impurities require to inject in assay specificity that how we can find out? and in Rs also how?
is it necessary to do HPLC calibration at wavelength 315nm if we are doing analysis at this wavelangth