Answer Posted / a.sudhakar
FT-IR IS CALIBRATED BY USING POLYSTYRENE FILM
PUT THE POLYSTYRENE FILM INTO THE SAMPLE HOLDER
AND SCAN IT FIXED WAVE NUMBERS ARE OBTAINED THEY ARE
3060
2849
1942
1601
1583
1154
1028
ACCEPTANCE CRITERIA IS +/- 1 CM
RESOLUTION: THE DIFFERENCE BETWEEN THE ABSORBANCE AT THE
ABSORBATION MINIMUM AT ABOUT 2870 CM AND THE ABSORBATION
MAXIMUM AT ABOUT 2849.5 CM SHALL BE MORE THAN 0.33
THE DIFFERENCE BETWEEN THE ABSORBANCE AT THE
ABSORBATION MINIMUM AT ABOUT 1589 CM AND THE ABSORBATION
MAXIMUM AT ABOUT 1583 CM SHALL BE MORE THAN 0.08.
| Is This Answer Correct ? | 29 Yes | 1 No |
Post New Answer View All Answers
What is column in chromatography?
if you have given one product then how you determine the impurity in that?
what is different when impergnated silica plates are used in separation of azo dyes using column chromatography?
for inorganic molecules require to do RS, Assay and disso?
what is the principle involved in turbidometry and nephelometry ?
why holium oxide prepared in 10% HCLO4 SOLUTION
CAN WE DETERMINE THE PEAK PURITY IN GAS CHROMATOGRAPHY ANALYSIS?
why lactose has less charges when packed in steel containers when compared with packing of lactose in polyamide material?
in DMF having extra impurities and in api COA also having extra imp than USP or BP product then how require to proceed?
Home heating oil or furnace oil what is made from? Can (base oil plus diesel/kerosen) = home heating oil?
How would you decide dissolution medium for NCE compound of class I drug
In performance Check of GC Why Hexadecane Peak is ConsideredÂ
Ph meter can show more than 14 ph reading? Why ph range in between 1 to 14 only?
If suppose 10 methods of dissolution given in pharmacopoeial for single content product then which method out of 10? or all 10 require to follow?
on saturation solubility study data how we can find out the bcs class of drug?