Golgappa.net | Golgappa.org | BagIndia.net | BodyIndia.Com | CabIndia.net | CarsBikes.net | CarsBikes.org | CashIndia.net | ConsumerIndia.net | CookingIndia.net | DataIndia.net | DealIndia.net | EmailIndia.net | FirstTablet.com | FirstTourist.com | ForsaleIndia.net | IndiaBody.Com | IndiaCab.net | IndiaCash.net | IndiaModel.net | KidForum.net | OfficeIndia.net | PaysIndia.com | RestaurantIndia.net | RestaurantsIndia.net | SaleForum.net | SellForum.net | SoldIndia.com | StarIndia.net | TomatoCab.com | TomatoCabs.com | TownIndia.com
Interested to Buy Any Domain ? << Click Here >> for more details...

How can v calibrate ftir ? Plz explain me

Answer Posted / a.sudhakar

FT-IR IS CALIBRATED BY USING POLYSTYRENE FILM

PUT THE POLYSTYRENE FILM INTO THE SAMPLE HOLDER
AND SCAN IT FIXED WAVE NUMBERS ARE OBTAINED THEY ARE
3060
2849
1942
1601
1583
1154
1028
ACCEPTANCE CRITERIA IS +/- 1 CM

RESOLUTION: THE DIFFERENCE BETWEEN THE ABSORBANCE AT THE
ABSORBATION MINIMUM AT ABOUT 2870 CM AND THE ABSORBATION
MAXIMUM AT ABOUT 2849.5 CM SHALL BE MORE THAN 0.33

THE DIFFERENCE BETWEEN THE ABSORBANCE AT THE
ABSORBATION MINIMUM AT ABOUT 1589 CM AND THE ABSORBATION
MAXIMUM AT ABOUT 1583 CM SHALL BE MORE THAN 0.08.

Is This Answer Correct ?    29 Yes 1 No



Post New Answer       View All Answers


Please Help Members By Posting Answers For Below Questions

what is mean by 40 in the dissolution basket mesh size?

714


As per ICH related substances stability trend limit from initial to shelf life

3182


We use Potassium Dichromate solution for the calibration of UV-Visible Spectrophotometer in UV region. My question is to calibrate visible region which solution could be used in photometric accuracy

1831


How do we fix the sample concentaryion in hplc method development. What is the basis?

1137


how to selecet an exact coloumn for an new molecule development by hplc how to select exact salt as buffer for new molecule development by hplc what is the the process to select the mode of saparation of compoundes by hplc what is the use of ph of buffer what is use of buffer,ph,organic phase,ans methods how the molecules get saparated in coloumn,

2528


how we can identify the impurity is coming below loq at transfering site?

808


What is the calibration of uv process and preparation

1936


In rs method development when we are going area normalization method to dilute standard method?

1473


analytical method validation require to with respect to release specification or shelf life specification?

732


Why are use silicon oil mr?

1150


I have compare C2H2-air and C2H2-N2O flame AAS on determination calcium. Both use same range of std to plot calibration curve. (2-6ppm) When i measure the sample with phosphate, KCl and LaCl, C2H2-N2O flame give false positive result, around 0.5ppm. When i measure the sample with phosphste, KCl and EDTA. C2H2-N2O flame also give 0.5ppm false positive. But both above mentioned sample would not give false positive when measured by C2H2-air flame. What is the reason?

3486


Why we select scan range from higer wavelength to lower wavelength in uv visible spectroscopy and ftir spectroscopy ? .

852


Principle of single pan analytical balance

5083


How require to develop GC method? how to select diluent, GAS, column selection and other chromatographic conditions?

683


AT WHAT CONDITIONS WILL YOU ANALYSE A SAMPLE,WHICH ARE PRESERVED AT OTHER THAN THE AMBIENT CONDITIONS?(ie COLD STORAGE SAMPLES BELOW 20* CENTIGRADE?

2012