acceptance criteria for ph meter, hplc
detector,pump,injector,oven when we calibrate.
Answer Posted / saptale maruti
Flow rate ± 0.2% of set value
Flow rate consistancy RSD NMT1.0%
Compositional Accuaracy : Difference in Absorbance is 0.01AU
Difference in time NMT 20SEc
Delay Volume NMT 4.5 (For 4 ml Delay Volume System)
Injection Volume Accuracy is : Average volume for 18
injection is 19.6 to 20.4µl
Linearity 0.999
Wavelenght Accuracy Maxima 273± 2 nm
Minima 205 and 245± 2 nm
Oven Temperature 2 degree from set temp.
For Ph ± 0.02
| Is This Answer Correct ? | 7 Yes | 2 No |
Post New Answer View All Answers
What is gelatinization?
is it nessesary to do solution stability for 7 days?
if tech transfer fails whay require to do?
In which situation we require to analytical method validation of excipient?
Why perchloric acid used for potentiometric titration
if identification threshold crosses the limits then what next step?
Why we select scan range from higer wavelength to lower wavelength in uv visible spectroscopy and ftir spectroscopy ? .
what is different when impergnated silica plates are used in separation of azo dyes using column chromatography?
how we can identify the impurity is coming below loq at transfering site?
Why only Copper standard is used to calibrate Atomic Absorption spectrophotometer?
what are the guidelines for analytical method validations?
which situation gc hs and gc als require to use?
analytical method validation require to with respect to release specification or shelf life specification?
how you start RS method development when for new product?
As per ICH related substances stability trend limit from initial to shelf life