acceptance criteria for ph meter, hplc
detector,pump,injector,oven when we calibrate.
Answer Posted / saptale maruti
Flow rate ± 0.2% of set value
Flow rate consistancy RSD NMT1.0%
Compositional Accuaracy : Difference in Absorbance is 0.01AU
Difference in time NMT 20SEc
Delay Volume NMT 4.5 (For 4 ml Delay Volume System)
Injection Volume Accuracy is : Average volume for 18
injection is 19.6 to 20.4µl
Linearity 0.999
Wavelenght Accuracy Maxima 273± 2 nm
Minima 205 and 245± 2 nm
Oven Temperature 2 degree from set temp.
For Ph ± 0.02
| Is This Answer Correct ? | 7 Yes | 2 No |
Post New Answer View All Answers
what is mean by extactable and leachable study?
how you determine moist.of NaOH sol or NaOH flacks
what is biorelivent dissolution media?
what you do when compound having dwel wavelength?
Colorimtry
How to set analyticl specification for combination products?
How to calculation of residual solvent methanol (3000 ppm) standard solution, and how many quantity used of test sample
How require to perform linearity as per ANVISA? What are the acceptance criteria?
how will you do the prep for unstable componds?
Identify problem faced when mass spectroscopy is used with HPLC system ?
What is mean by PDR?
we are performed SOR for a particular product the limit is -6 to -10.We face particular bathes it is not meeting the specification.we performed diffrent instrument diffrent analyst we are getting diffrent diffrent results.solvent is dimethyl sulfoxide what could be the reasion
What is aggregate and fragments in SEC?
What is diffrence between extractable volume and deliverable volume? Answer pls
How to start the dissolution development for unknown tab?