How did you conclude that when stressing a compond, all
degradation products have eluted in chromatogram?
Answer Posted / s.elumalai
The main aim of degradation study is to separate know
impurities in the case when we do not have know impurities
or to generate impurities from the sample by degradation.
Obviously one should know the process related impurities or
the degradation pathways. From peak purity by PDA or Mass
analysis one can conclude the purity of each peak eluting,
beyond that mass balance also account for peaks.If mass is
not balanced then one can conclude that some of the compound
are not eluted or may not have UV absorption.
| Is This Answer Correct ? | 7 Yes | 0 No |
Post New Answer View All Answers
how to decide for one product require water content or LOD?
Why sometime potassium ortho phosphate mix with acetone use for sample preparation for hplc
In HPLC calibration, caffeine is used as primary standard for wave length calibration due to caffeine is having dual maxima at 273 & 205 nm and one minima at 245 nm. Any body can give reference of these details from any pharmacopeia (with chapter no.) or any other authentic guideline?
if you given one product then which tests you will perform?
UV and PDA detector, which have less signal to noise ratio?
how a particular wavelength can be different for a particular compund while analysing by uv and by HPLC.
why should we perform dissolution PVT calibration only by UV spectrometer, not by HPLC ?
WHICH STANDARD WE USE IN SILICA ANALYSIS IN SILICA-GEL METHOD FOR SOLID SAMPLE
which are the sizes of capsules?
Why we check moisture,ash & AIA in product
Basic principle of ESI?
Why to use propyl paraben in HPLC gradient calibration
Which are the diffrent grades of api in pharma?
cefoperozone and sulbactam inj. hplc test method
iam usig ph buffers merk. manually how to prepare ?