Golgappa.net | Golgappa.org | BagIndia.net | BodyIndia.Com | CabIndia.net | CarsBikes.net | CarsBikes.org | CashIndia.net | ConsumerIndia.net | CookingIndia.net | DataIndia.net | DealIndia.net | EmailIndia.net | FirstTablet.com | FirstTourist.com | ForsaleIndia.net | IndiaBody.Com | IndiaCab.net | IndiaCash.net | IndiaModel.net | KidForum.net | OfficeIndia.net | PaysIndia.com | RestaurantIndia.net | RestaurantsIndia.net | SaleForum.net | SellForum.net | SoldIndia.com | StarIndia.net | TomatoCab.com | TomatoCabs.com | TownIndia.com
Interested to Buy Any Domain ? << Click Here >> for more details...

control of resolution in calibration of IR

Answer Posted / nirav intas pharma

control of resolution in IR:

Take spectrum of polystyrine film,

the abs minima at 2870 cm- & abs maxima at 2849.5cm- is
greater than 0.33

the abs minima at 1589 cm- & abs maxima at 1583 cm- is
greater than 0.08

Is This Answer Correct ?    12 Yes 0 No



Post New Answer       View All Answers


Please Help Members By Posting Answers For Below Questions

UV and PDA detector, which have less signal to noise ratio?

1177


2. Two grams of Benzoic acid are dissolved in 200 ml of water and extracted with 200 ml of diethyl ether. The distribution coefficient of benzoic acid is 100, and its dissociation constant is 6.5 10-5. Calculate the distribution ratio (D) of benzoic acid at pH 2, 5, and 6. 3. Calculate D at pH 2 to 10 (1 unit apart) in the above problem, and plot D versus pH.

2263


if instrument calibration fails then what require to do?

1018


How can I develope method of dissolution by HPLC OR UV

2456


Why are use silicon oil mr?

1258


we can use expired sample for validation and analyst qualification?

862


What is column in chromatography?

2290


how a particular wavelength can be different for a particular compund while analysing by uv and by HPLC.

2364


how you start RS method development when for new product?

779


What is the diference between residual solvents and organic volatile matter

2110


What is stray light in uv what are units for stray light

1720


why we should take dst factor for below 1%moisture samples

2716


in which situation require to take incident in validation?

1783


How to calculate coreletion coefficient

1765


How to calculation of residual solvent methanol (3000 ppm) standard solution, and how many quantity used of test sample

3330