control of resolution in calibration of IR

Answer Posted / nirav intas pharma

control of resolution in IR:

Take spectrum of polystyrine film,

the abs minima at 2870 cm- & abs maxima at 2849.5cm- is
greater than 0.33

the abs minima at 1589 cm- & abs maxima at 1583 cm- is
greater than 0.08

Is This Answer Correct ?    12 Yes 0 No



Post New Answer       View All Answers


Please Help Members By Posting Answers For Below Questions

how to select short coloum and long coloumn for new molecule

1817


How to calculation of residual solvent methanol (3000 ppm) standard solution, and how many quantity used of test sample

2529


In Dissolution Test why limit is define Q+5% what is the role of +5%.

2569


WHAT IS THE CRITERIA FOR SELECTION OF TIME OF DISSOLUTION AND THE MEDIUM OF DISSOLUTION?

2139


why xterra column require to use at higher ph?

431






how we can identify the impurity is coming below loq at transfering site?

382


2. Two grams of Benzoic acid are dissolved in 200 ml of water and extracted with 200 ml of diethyl ether. The distribution coefficient of benzoic acid is 100, and its dissociation constant is 6.5 10-5. Calculate the distribution ratio (D) of benzoic acid at pH 2, 5, and 6. 3. Calculate D at pH 2 to 10 (1 unit apart) in the above problem, and plot D versus pH.

1760


in DMF having extra impurities and in api COA also having extra imp than USP or BP product then how require to proceed?

346


what is the principle of UV Vis spectroscopy, AAS, ICP OES,ICPAES, ICP-MS and FTIR

8428


What is diffrence between extractable volume and deliverable volume? Answer pls

3928


If inhouse hplc related substance method is completly diffrent from Usp for finished proďuct with diffrent impurities then how require to prove method equivalecy?

308


Why a1% value is used for some product ? What is the criteria for selection a1% value ?

1863


in which situation require to change rs specification?

272


Colorimtry

1169


how many types of balance calibration test?

1721