What is calibration procedure of gredient HPLC pump
Answer Posted / prerna
Gradient Programme for port A and C :
00 : 100
00 : 100
25 : 75
25 : 75
50 : 50
50 : 50
75 : 25
75 : 25
100 : 00
100 : 00
Above programme also run for port B and D
An increasing stepwise piramid will appear. Integrate that
peak and also split each step of the peak.
Find out its height and area of height
By using the formula ( which I don't remeber ), mean flow
of solvent at each step.
Flow : Limit
25 (24-26)
50 (49-51)
75 (74-76)
100 (100)
If the observed value is within the acceptance limit then
that pump passes the test.
| Is This Answer Correct ? | 10 Yes | 10 No |
Post New Answer View All Answers
Ph meter can show more than 14 ph reading? Why ph range in between 1 to 14 only?
what is third integration
From where require to take the RLD sample?
what is mean by nitrosamine impurities?
Please tell me about the pH of Polycaboxylic ether is it in the 5-6 range ever or more than 6
how can give the expiry period and restadardisation of volumetric solution
How require to perform linearity as per ANVISA? What are the acceptance criteria?
How would you decide dissolution medium for NCE compound of class I drug
why salisylic acid not using now days for disdolution calibration?
Describe your field experience sample type collected,sample techniques,field measurements taken and equipment used?
which are the sizes of capsules?
How to determine water content of bis tetrazole amine mono ammonium salt
what are the guidelines for analytical method validations?
sop of a uv visible spectrophotometer double beam elico model
Ratio of Water and Acetonitrile(diluent) is not recommended to use in cleaning method validation,why?