Golgappa.net | Golgappa.org | BagIndia.net | BodyIndia.Com | CabIndia.net | CarsBikes.net | CarsBikes.org | CashIndia.net | ConsumerIndia.net | CookingIndia.net | DataIndia.net | DealIndia.net | EmailIndia.net | FirstTablet.com | FirstTourist.com | ForsaleIndia.net | IndiaBody.Com | IndiaCab.net | IndiaCash.net | IndiaModel.net | KidForum.net | OfficeIndia.net | PaysIndia.com | RestaurantIndia.net | RestaurantsIndia.net | SaleForum.net | SellForum.net | SoldIndia.com | StarIndia.net | TomatoCab.com | TomatoCabs.com | TownIndia.com
Interested to Buy Any Domain ? << Click Here >> for more details...

Why we use caffeine for HPLC calibration in Detector wave
ength accuracy test

Answer Posted / devendrakumar

1. caffeine easily available in market.
2. In general we all use caffeine as a standard for
calibration of HPLC specially for (Calibration of Wave
Length Accuracy) since it will scan the whole of the UV
range i.e It gives a multiwavelength response like
first maxima at 205 nm
second maxima at 273 nm &
minima at 245nm
3. It is very stable, we can use a long time for analysis
i.e. it is not degradable
4. In this case we can use Erbium Perchlorate which will
scan the whole range(200 nm to 600 nm).It is stable and
little costly but not easily available in market.


d.devendrakumar

Is This Answer Correct ?    100 Yes 5 No



Post New Answer       View All Answers


Please Help Members By Posting Answers For Below Questions

identification is for unknown? qualification for known? reporting for LOQ?

780


How to set analytical specification for combination products?

876


How do we quantify crystaline and amarpous forms by using (NMR, XRD)spectroscopic techniques? Which any others instruments are useful for this quantification? explain

2440


Why we use potassium dichromate in uv calibration Exact reason behind it??

1326


analytical method validation require to with respect to release specification or shelf life specification?

851


How require to perform linearity as per ANVISA? What are the acceptance criteria?

1401


For limit test of heavy metals in BP, Method C require that the substance is ignited at a temperature not exceeding 800 °C. Why confines such the temperature?

4135


why multimedia dissolution require to do?

867


on saturation solubility study data how we can find out the bcs class of drug?

956


How do we get end points and how many end points are possible for citric acid and di-acid not theorotically answer should be given practically.

2631


why octanol used to determine the partition coefficient ?

2499


for which product require to do content uniformity? what is limit of cu?

878


how will you do the prep for unstable componds?

2070


what is %labelled amount in content uniformity of dosage unit and its calculation?

5303


in monograph 7 imp are there but in vendor showing 5 imp do not req to include 2 imp for that what will be justification?

786