how to pepare 6.8 pH phosphate buffer

Answer Posted / thamer

For Phosphate Buffer pH 6.8, Mixed :: Dissolve 28.80 g of disodium hydrogen orthophosphate and 11.45 g of potassium dihydrogen orthophosphate in sufficient water to produce 1000 ml.
<or>

Citro-phosphate Buffer pH 6.8
Citro-phosphate Buffer pH 6.8 Phosphate buffer solution pH 6.8
Mix 77.3 ml of a 7.15% w/v solution of disodium hydrogen orthophosphate with 22.7 ml of a 2.1% w/v solution of citric acid .

Is This Answer Correct ?    3 Yes 0 No



Post New Answer       View All Answers


Please Help Members By Posting Answers For Below Questions

In GC calibration,what is the procedure to calibrate the TCD (thermal counductivity detector)suggest ?

2720


Colorimtry

1182


what is the principle of UV Vis spectroscopy, AAS, ICP OES,ICPAES, ICP-MS and FTIR

8469


what is the structure of DOWEX-50

4931


how to qualify the impurity?

328






Why use only methanol,used determinationof moisture content not use other solvents

889


AT WHAT CONDITIONS WILL YOU ANALYSE A SAMPLE,WHICH ARE PRESERVED AT OTHER THAN THE AMBIENT CONDITIONS?(ie COLD STORAGE SAMPLES BELOW 20* CENTIGRADE?

1614


How doing qbd practically?

434


in which situation require to change rs specification?

278


inhouse product is in capsule form in combination and RLD is in tablet form then can we proceed for multimedia CDP? in inhouse capsule product disso is paddle with sinker in release media is there then RLD product in tablet form then with same as paddle with sinker we can proceed n.a.?

298


What is the difference between chromatographic purity and related substances?

796


How to decide assay range for non pharmacopeial API analysis by HPLC? Could you give me any reference for same e.g Guidelines or Paper publications?

2875


I have compare C2H2-air and C2H2-N2O flame AAS on determination calcium. Both use same range of std to plot calibration curve. (2-6ppm) When i measure the sample with phosphate, KCl and LaCl, C2H2-N2O flame give false positive result, around 0.5ppm. When i measure the sample with phosphste, KCl and EDTA. C2H2-N2O flame also give 0.5ppm false positive. But both above mentioned sample would not give false positive when measured by C2H2-air flame. What is the reason?

2985


How to know the estimated LOQ concentration in ppm

739


we are performed SOR for a particular product the limit is -6 to -10.We face particular bathes it is not meeting the specification.we performed diffrent instrument diffrent analyst we are getting diffrent diffrent results.solvent is dimethyl sulfoxide what could be the reasion

1362