How to prepare 50 ppm ammonia solution form 25% ammonia
solution?
Answer Posted / greg
Convert % to ppm. This is probably faster.
Ammonia Solution
1% = 10000ppm
25% = 250000 (25/100 = 0.25 X 1000000)
C1V1 = C2V2
C1 = 50 ppm (target solution concentration)
V1 = 100 mL (targer solution volume)
C2 = 250000 ppm (stock solution concentration)
V2 = X (stock solution volume required)
V2 = 0.02 mL diluted to 100mL in a volumetric flask
Is This Answer Correct ? | 8 Yes | 2 No |
Post New Answer View All Answers
if your impurity coeluting with each other in that situation how require to set specification? is it acceptable?
Why only Copper standard is used to calibrate Atomic Absorption spectrophotometer?
WHAT IS THE PARAMETER FOR THE SELECTION OF MOBILE PHASE IN HPLC?
process of Diclofenac sodium,IP.
What is the diffrence in japan mkt requirement in analytical method validation over US?
how to selecet an exact coloumn for an new molecule development by hplc how to select exact salt as buffer for new molecule development by hplc what is the the process to select the mode of saparation of compoundes by hplc what is the use of ph of buffer what is use of buffer,ph,organic phase,ans methods how the molecules get saparated in coloumn,
Ph meter can show more than 14 ph reading? Why ph range in between 1 to 14 only?
In HPLC calibration, caffeine is used as primary standard for wave length calibration due to caffeine is having dual maxima at 273 & 205 nm and one minima at 245 nm. Any body can give reference of these details from any pharmacopeia (with chapter no.) or any other authentic guideline?
why we are using hexane in calibration of number of drop per mL
Is it nessesary all multimedia dissolution require descriminatory?
How we performed the force degradation for drug substance, is any specific guideline is available for each parameter(Acidic, basic, oxidation,heat)? what conditions you mentained for above parameters.
If change in specification which parameter require to do for validation? if change in chromtographic condition then which parameter? if api change then which parameter? if change composition then which parameter? if old method not work out then whicj parameter? if additional one impurity added then which parameter of validation require to do on above each conditions? elaborate separately
How much is the standard area for glc analysis
Why we check moisture,ash & AIA in product
what is %labelled amount in content uniformity of dosage unit and its calculation?