In the test for control of absorbance in UV calibration why
do we use only potassium dichromate and what is purpose of
taking a specified amount of 57.0-63.0mg?
Answer Posted / kishor solanke (alkem)
In UV specrophotometer calibration during control of
absorbance paramater the K2Cr2O7 is taken 60mg±3mg because
the resulting solution gives the absorbance between 0.2 to
0.9 units which is the high performance & optimum range for
UV spectrophotometer.
| Is This Answer Correct ? | 14 Yes | 2 No |
Post New Answer View All Answers
Tell me about analytical method validation in QC
What is the principle of HPLC, GC, LCMS, GCMS,LC QQQ, GC QQQ, LC Q TOF and GC Q TOF. What are the applications and Specificity?
How they found 1mL of K.F reagent is equivalent to 5mg of water and if we change the composition of K.F reagent, is it can neutralize more amount of water?
In which situation we require to prepare the standard solution from sample in Related substance method?
How to set analyticl specification for combination products?
how to calibrate hplc & gc
in which situation require to take incident in validation?
what are the standard methods for HPLC?
What is the difference between spectro meter and spectro photo meter?
Difference between hlaf and rlaf
what is mean by 40 in the dissolution basket mesh size?
What is the difference between Discriminating media and DPDM(Dissolution Profile with Different Media)
What should be the level/ Percentage of Arsenic in Arsenic free zinc.
please explian me about area normalisation method,diluted standard method,impurity vs impurity and which one has to be selected in method development
What is split ratio in Gc? Splitless? how requirr to select?