In the test for control of absorbance in UV calibration why
do we use only potassium dichromate and what is purpose of
taking a specified amount of 57.0-63.0mg?
Answer Posted / kishor solanke (alkem)
In UV specrophotometer calibration during control of
absorbance paramater the K2Cr2O7 is taken 60mg±3mg because
the resulting solution gives the absorbance between 0.2 to
0.9 units which is the high performance & optimum range for
UV spectrophotometer.
| Is This Answer Correct ? | 14 Yes | 2 No |
Post New Answer View All Answers
if you have given a blank water how you analyse by general analytical techniques?
Which parameters require to do in tech transfer?
How to choose the which salt is suitable for mobilephase
why lactose has less charges when packed in steel containers when compared with packing of lactose in polyamide material?
How we can identify process related and degradation impurity in single method with short period?
If combination product how require to identify which imp is of which api?
how to calculate elemental impurities?
What is the formula for relative diffrence for standard solution in solution stability in validation?
Why six unit used for precision?
can we use the same detector in HPLC as well GC and what could be the differences we can find in the final chromato graph in any aspects?
WHAT IS THE PURPOSE WE USE HCL, CHOPPER, BURNER, MONOCHROMATOR in AAS? PLEASE EXPLAIN BRIEFLY
How to calculation of residual solvent methanol (3000 ppm) standard solution, and how many quantity used of test sample
how many types of balance calibration test?
what parameters we will consider while developing a HPLC method and how we confirm our HPLC method is valid?
acceptance criteria for lod & loq by standard deviation of response and slope??