Golgappa.net | Golgappa.org | BagIndia.net | BodyIndia.Com | CabIndia.net | CarsBikes.net | CarsBikes.org | CashIndia.net | ConsumerIndia.net | CookingIndia.net | DataIndia.net | DealIndia.net | EmailIndia.net | FirstTablet.com | FirstTourist.com | ForsaleIndia.net | IndiaBody.Com | IndiaCab.net | IndiaCash.net | IndiaModel.net | KidForum.net | OfficeIndia.net | PaysIndia.com | RestaurantIndia.net | RestaurantsIndia.net | SaleForum.net | SellForum.net | SoldIndia.com | StarIndia.net | TomatoCab.com | TomatoCabs.com | TownIndia.com
Interested to Buy Any Domain ? << Click Here >> for more details...

same sample on two preapration gives differnt result during
hplc analysis

Answer Posted / reet

there may be any one of the reasons:
1. Air bubble in the injector needle (needle wash 2-3 times0
2. carry over might be there.
3. random variation in the flow rate
4. Integration problem (chk the processing method)
5. Sample itself may a problem !
i. solubility issues
ii.sample(generally proteins may be sticking to the
walls of the sample vial and also may be getting
irreversibly stuck on the surface of the column)
iii. temperature change.

Is This Answer Correct ?    5 Yes 0 No



Post New Answer       View All Answers


Please Help Members By Posting Answers For Below Questions

If inhouse hplc related substance method is completly diffrent from Usp for finished proďuct with diffrent impurities then how require to prove method equivalecy?

735


about method developement in hplc

1667


What is the accrptance criteria in RSD for RS method precision on basis of impurity percentages?

770


what you do when compound having dwel wavelength?

704


How to choose the which salt is suitable for mobilephase

3178


How would you decide dissolution medium for NCE compound of class I drug

2196


WHAT IS THE USE OF STARTER ?????????

2357


how you start the solubility study?

770


what is the procedure for cleaning of lenses of hatr accesory of ftir instrument?

2231


what is procedure of actual procedure for registration of drug?

1959


My question about gas chromatography sulfur chemiluminsecence detector. I test unknown sample gas by GC-SCD (calibrated ) and the result of *H2S is 279 PPM , *but when I test the same sample with the GC-TCD (calibrated ) the value of *H2S is 2500 PPM . I'd like to inform you that both GCs are calibrated and have very good operation conditions with stable parameters . the question is if the sample gas with higher H2S over detection limits of SCD detector (1000 ppm). why the result it 279 ppm Best regards

1549


Why only Copper standard is used to calibrate Atomic Absorption spectrophotometer?

2671


what is %labelled amount in content uniformity of dosage unit and its calculation?

5138


what is the acceptance criteria for enteric coated tablets in 0.1n hcl validation in each parameter?

836


In performance Check of GC Why Hexadecane Peak is Considered 

2989