In case of HPLC coloums what is the difference Between C8 &
C18 coloums & how can we select difference type of coloums
like Inertial ODS , Hypersil BDS , ODS etc.?
Answer Posted / paresh jobanputtra, piramal(ah
in case of C8 column,C8 carbon chain is attached to the packing material
and in case of C18(ODS) column, C18 carbon chain is attached to the packeting material.
eg. OH
ยก
OH-SI-O-C8 orC18
!
OH
Some names of specially treated stationary phases:suitable for separation of BASIC compds.
like Inertsil:"Inert towards bases"
BDS :"Base deactivated silica"
different make columns have different characteristics
like different Carbon load,Specific pore volume(empty spaces betn particles),Specific surface area,Ligand density etc.
If the carbon load is high it means specific pore volume is lesss so retained capability of packing material is high so capacity factor is high. so retemtion is better of this column.
and selection of column is depends on nature of product, and no of trials in developement.
Regards,
Paresh Jobanputtra.
| Is This Answer Correct ? | 26 Yes | 3 No |
Post New Answer View All Answers
HI,I CLEARED BOB CLERK EXAM. MY INTERVIEW WILL BE ON 9TH OCTOBER,2010.PLEASE SEND ME INTERVIEW QUESTIONS AND ANSWERS. THANK YOU.
in api coa contains only process impurities and in product coa contains degradation impurities?
What is gelatinization?
From where require to take the RLD sample?
Which type of column should i use to check the purity of high molecular weight protein using HPLC reverse phase column chromatography? Hi everyone. I wanna to check the purity of high molecular weight protein (collagen) with MW of ~130 kDa using a HPLC. I know C18
As per ICH related substances stability trend limit from initial to shelf life
Is it nessesary all multimedia dissolution require descriminatory?
how decide the clining method and cleaning method validation require for this perticular products?
What are usp limits for theoritical plats,resoution,tailing factor,peak to valley ratio
IS THERE ANY EQUIPMENT TO CHECK AND CALCULATE THE POLARITY OF A LIQUID?
what is difference in ods I ods II is this is realy from sorbent / packing characteristics or from other means
Why we select scan range from higer wavelength to lower wavelength in uv visible spectroscopy and ftir spectroscopy ? .
What is delay volume?
pharma industries oriented
What is the calibration of uv process and preparation