Golgappa.net | Golgappa.org | BagIndia.net | BodyIndia.Com | CabIndia.net | CarsBikes.net | CarsBikes.org | CashIndia.net | ConsumerIndia.net | CookingIndia.net | DataIndia.net | DealIndia.net | EmailIndia.net | FirstTablet.com | FirstTourist.com | ForsaleIndia.net | IndiaBody.Com | IndiaCab.net | IndiaCash.net | IndiaModel.net | KidForum.net | OfficeIndia.net | PaysIndia.com | RestaurantIndia.net | RestaurantsIndia.net | SaleForum.net | SellForum.net | SoldIndia.com | StarIndia.net | TomatoCab.com | TomatoCabs.com | TownIndia.com
Interested to Buy Any Domain ? << Click Here >> for more details...

How can we calibrate the burette dispencing capacity in
metrorm kf titration calibration

Answer Posted / subash

First thing is that clean the burette about 3 to 4 times
with disttled water for removal of air gaps or previous
solution.
After sufficient cleaning fill the bureete with distilled
water and make sure that there is no bubbles formation this
lead to misreading.
First weigh the required volumetric flask similar to the
didpensing burette volume using balance i.e initial weight
Dispense the disttled water completely in to volumetric
flask and weigh the volumetric flask again i.e Final weight

Actual weight= Final weight-Initial weight

The weight can be converted to Volume using formula

Density=mass/volume

Density changes with temperature and note the down the
temperature at which calibration is done.

Is This Answer Correct ?    4 Yes 1 No



Post New Answer       View All Answers


Please Help Members By Posting Answers For Below Questions

In HPLC Calibration, On which basis RSD Limit of noise test is fixed (NMT 33.0 % )

3452


acceptance criteria for lod & loq by standard deviation of response and slope??

2559


USP methodology, EP methodology, IP methodology, among three if possible to use one methodology for qualify working standard to use USP, EP, IP ? Please explain...

1989


What is the acceptance criteria for moisture balance when calibrated with sodium sulfate

2014


UV and PDA detector, which have less signal to noise ratio?

1105


What is aggregate and fragments in SEC?

2081


what is mean by ambient temperature?

893


what is mean by 40 in the dissolution basket mesh size?

741


why multimedia dissolution require to do?

761


Why do we use KMnO4 in the test of control of obsorbance ? and why do we take specific quantity i.e 57-63mg?

2799


What is control room temperature and which guide line says?

1604


How require to perform linearity as per ANVISA? What are the acceptance criteria?

1299


WHY pKa value is more important than pH value of the mobile phase in HPLC?

1741


What is the diffrence in japan mkt requirement in analytical method validation over US?

699


before starting analytical method valodation what you checking? and how giving preference to start validation?

765