What is discriminating media? How can u justify/select that
which media (pH etc...)is suitable for your dissolution study
of your compound?
Answer Posted / bhupal24
A releasing medium is the one that promotes
solubilization of the drug substance from a pharmaceutical
dosage form without further qualification.
When selecting the dissolution medium, physical and
chemical data for the drug substance and drug product need
to be considered,selecting the composition of the medium,
the influence of buffers, molarity, pH, and surfactants
on the solubility and stability of the drug also need to be
evaluated,
The most common dissolution medium is dilute
hydrochloric acid, however other media commonly used
includes buffers in the physiologic pH range of 1.2 to 7.5,
simulated gastric or intestinal fluid (with or without
enzymes), water, and surfactants (with or without acids or
buffers)such as polysorbate 80, sodium lauryl sulfate, and
bile salts. The use of aqueous-organic solvent mixtures,
while generally discouraged, can also be used if justified.
Enzymes are also sometimes used in the media when testing
gelatin capsule products.
Media volumes are typically in the range of 500-1000
mL, with 900 mL the most common volume. Volumes as high as
2-4 L have been used, and as low as 100 mL for high potency
(low dosage strength) drug formulations.
| Is This Answer Correct ? | 9 Yes | 2 No |
Post New Answer View All Answers
what is turbidimetric titration?Give 2 examples.
What is the difference between chromatographic purity and related substances?
How to know the estimated LOQ concentration in ppm
Ratio of Water and Acetonitrile(diluent) is not recommended to use in cleaning method validation,why?
What is the calibration of uv process and preparation
Why perchloric acid used for potentiometric titration
why glutent are detected in the rice cereal baby food product even manufacturer claimed that they are using rice and milk only?we have using ELISA to do the test,and rice supposed not containing any glutent,rite?We already repeat the test so many times and it still detected.just wondering where the glutent came from?
how you fix the limits of impurities?
what is partion and column chromatography
is it necessary to do HPLC calibration at wavelength 315nm if we are doing analysis at this wavelangth
USP methodology, EP methodology, IP methodology, among three if possible to use one methodology for qualify working standard to use USP, EP, IP ? Please explain...
can we use the same detector in HPLC as well GC and what could be the differences we can find in the final chromato graph in any aspects?
what is the difference between potentiometric titration and karl fischer titration?
why we are using hexane in calibration of number of drop per mL
How to calculation of residual solvent methanol (3000 ppm) standard solution, and how many quantity used of test sample