Golgappa.net | Golgappa.org | BagIndia.net | BodyIndia.Com | CabIndia.net | CarsBikes.net | CarsBikes.org | CashIndia.net | ConsumerIndia.net | CookingIndia.net | DataIndia.net | DealIndia.net | EmailIndia.net | FirstTablet.com | FirstTourist.com | ForsaleIndia.net | IndiaBody.Com | IndiaCab.net | IndiaCash.net | IndiaModel.net | KidForum.net | OfficeIndia.net | PaysIndia.com | RestaurantIndia.net | RestaurantsIndia.net | SaleForum.net | SellForum.net | SoldIndia.com | StarIndia.net | TomatoCab.com | TomatoCabs.com | TownIndia.com
Interested to Buy Any Domain ? << Click Here >> for more details...

How to calculate LOD & LOQ in HPLC Validation Method

Answer Posted / sunil.madhukar rao kulkarni

We can calculate LOD and LOQ values using linearity
graph,insted of sigma value STEYX can be considered for
more acuurate vale of LOD and LOQ

where STEYX : Standard error of the pridected Y-value for
each X in a regression

Final calulation given as below :

LOD = STEYX x 3/ Slope

LOQ = STEYX x 10 /Slope

Is This Answer Correct ?    44 Yes 10 No



Post New Answer       View All Answers


Please Help Members By Posting Answers For Below Questions

How to set analytical specification for combination products?

774


Which parameters require to do in tech transfer?

681


For titration in anhydrous media with perchloric acide, if lack of titrator, Which indicator is been used for replacement. How calculate pH of test solution to choose suitable indicator?

2262


i want model written test paper for the post of analyst in IOCL(indian oil corporation ltd) any one pls help me

2669


What is viod volume and peak purity in HPLC?

2811


Tell me something about Vitamin A test method by HPLC

2627


How can we confirm the HPLC column is end-capped or not? Is it possible to identify by physical appearance?

3154


Why acetonitrile and water are used as extraction solvent when analysis melamine? I thought they are miscible and won't be able to separate...

1632


Ratio of Water and Acetonitrile(diluent) is not recommended to use in cleaning method validation,why?

2896


using gradient pressure in gas chromatography are not ?using gradient pressure why

3333


If change in specification which parameter require to do for validation? if change in chromtographic condition then which parameter? if api change then which parameter? if change composition then which parameter? if old method not work out then whicj parameter? if additional one impurity added then which parameter of validation require to do on above each conditions? elaborate separately

796


from where i get mortar pastle for glass bottle crush? it's required for testing of it?

2224


How can we calculate offset value pH measurment

1463


What is column in chromatography?

2190


How doing qbd practically?

916