How can we calibrate THe Auto Titrator ?

Answer Posted / krishnakumar

response of electrode- using ph 4.0 7.0 and 9.0 read the
millivolt reading and make a graph with ph vs mv .graph
slope is between 90 percent to 110percent

burette . by selecting 1 ml 2ml 5 ml 10 ml and 20 ml
titrant is water. check the weight of water in each step
and 1 percent limit is allowed from the selected volume

temperature- compare with calibrated thermometer at minimum
4 intervals .1 percent error is allowed from the calibrater
thermometer.correlation coefficient of temperature graph
actual and set value should be greater than 0.9

Is This Answer Correct ?    12 Yes 6 No



Post New Answer       View All Answers


Please Help Members By Posting Answers For Below Questions

What is electrode potential? : Electrochemistry Corrosion

593


What is the principle of Particle size analysis and zeta potential

2042


A polymer's structure influences its physical properties. Describe two structural variations that are possible for polypropylene but not for polyethylene. : Polymer Chemistry

611


Explain the dipole moment direction for methanol?

645


What are the factors which affect corrosion? : Electrochemistry Corrosion

537






What is the process for the removal of salts from brackish water or seawater to make it usable?

587


Explain how do you prepare a solution of 1 m hcl?

582


How tripod is used?

630


Why are olefins (alkenes) good monomers for polymerization reactions? : Polymer Chemistry

896


Explain the use of glacial acetic acid?

657


Explain what is an oxidizing agent?

622


How would you describe the orientation of the ligands around the zr center in the homogeneous zirconium catalyst? : Polymer Chemistry

556


What is the number of protons in an atom?

573


Explain the difference in the modern periodic table and mendeleevs table?

565


Define a substituted hydrocarbon?

617