How can we calibrate THe Auto Titrator ?
Answer Posted / krishnakumar
response of electrode- using ph 4.0 7.0 and 9.0 read the
millivolt reading and make a graph with ph vs mv .graph
slope is between 90 percent to 110percent
burette . by selecting 1 ml 2ml 5 ml 10 ml and 20 ml
titrant is water. check the weight of water in each step
and 1 percent limit is allowed from the selected volume
temperature- compare with calibrated thermometer at minimum
4 intervals .1 percent error is allowed from the calibrater
thermometer.correlation coefficient of temperature graph
actual and set value should be greater than 0.9
| Is This Answer Correct ? | 12 Yes | 6 No |
Post New Answer View All Answers
Explain the difference between hcl acid and hcl gas?
What is the principle of Particle size analysis and zeta potential
Define an origin of electrode potential? : Electrochemistry Corrosion
How are atoms of the isotopes different from one another? Two isotopes of carbon are carbon-12 and carbon-14.
Define dipole moment?
How to convert -40 of into centigrade?
Explain how do you separate the colors of ink?
Explain what is an oxidizing agent?
Define a triglyceride?
Explain the dipole moment direction for methanol?
What is acid anhydride of h2co3?
Observe the following number to four significant figures and express the result in standard exponential notation: 0.006543210?
Suppose u have combipack product and u get 2 or more peaks which are other than there placebo then what we can do
can you email me because I have a report and I need you to answer these questions if you would like to if not I completely under stand.
what are the role of chemical education in sustainable development