how peak can be sharp in hplc when it is come flat
Answer Posted / vivek k
Sample solvent stronger than the mobile phase Dissolve
sample in mobile phase or at least reduce the strength of
the sample solvent as much as possible
Sample mass overload Reduce the amount (mass) of sample
injected.
Silanol interaction with amines • Reduce mobile phase pH to
< 3.0
• Increase mobile phase ionic strength.
25mM to 50mM recommended
• Add a competing amine to the mobile phase.
10 mM TEA is usually sufficient.
• Select a stationary phase with a lower silanol activity.
Adsorption of acids on silica • Increase salt
concentration in the mobile phase
25 mM to 50 mM is usually sufficient
• Reduce the pH of the mobile phase to < 3.0
• Add a competing organic acid.
1% acetic acid or 0.1% TFA is usually sufficient
Column void Replace the HPLC column.
Attempts to fill-in the void are seldom worth the effort.
| Is This Answer Correct ? | 10 Yes | 0 No |
Post New Answer View All Answers
If inhouse hplc related substance method is completly diffrent from Usp for finished proďuct with diffrent impurities then how require to prove method equivalecy?
how to selecet an exact coloumn for an new molecule development by hplc how to select exact salt as buffer for new molecule development by hplc what is the the process to select the mode of saparation of compoundes by hplc what is the use of ph of buffer what is use of buffer,ph,organic phase,ans methods how the molecules get saparated in coloumn,
what is the procedure for cleaning of lenses of hatr accesory of ftir instrument?
why multimedia dissolution require to do?
While performing TOC sst analysis Zero shift disabled & sample analysis zero shift enabled why?
In HPLC (Waters) software Empower1,2 &3 what is the updation for each version
how you determine moist.of NaOH sol or NaOH flacks
Which are the diffrent grades of api in pharma?
How to decide assay range for non pharmacopeial API analysis by HPLC? Could you give me any reference for same e.g Guidelines or Paper publications?
Why perchloric acid used for potentiometric titration
What is shaking level in GC?
How to calibration of the uv spectroscopy and its test?
in which situation require to change rs specification?
cefoperozone and sulbactam inj. hplc test method
How can I develope method of dissolution by HPLC OR UV