what is karl fischer principle and reaction ,describe it
Answer Posted / suresh babu
The fundamental principle behind it is based on the Bunsen
reaction between iodine and sulfur dioxide in an aqueous
medium.
Water and iodine are consumed in a 1:1 ratio in the above
reaction. Once all of the water present is consumed, the
presence of excess iodine is detected voltametrically by
the titrator’s indicator electrode. That signals the end-
point of the titration.
The amount of water present in the sample is calculated
based on the concentration of iodine in the Karl Fisher
titrating reagent (i.e., titer) and the amount of Karl
Fisher reagent consumed in the titration.
B·I2 + B·SO2 + B + H2O → 2BH+I− + BSO3
BSO3 + ROH → BH+ROSO3−
Is This Answer Correct ? | 28 Yes | 3 No |
Post New Answer View All Answers
we are performed SOR for a particular product the limit is -6 to -10.We face particular bathes it is not meeting the specification.we performed diffrent instrument diffrent analyst we are getting diffrent diffrent results.solvent is dimethyl sulfoxide what could be the reasion
In Assy by potentio metry titration method how select MET U ,DET U, SET U,
WHAT IS THE USE OF STARTER ?????????
My question about gas chromatography sulfur chemiluminsecence detector. I test unknown sample gas by GC-SCD (calibrated ) and the result of *H2S is 279 PPM , *but when I test the same sample with the GC-TCD (calibrated ) the value of *H2S is 2500 PPM . I'd like to inform you that both GCs are calibrated and have very good operation conditions with stable parameters . the question is if the sample gas with higher H2S over detection limits of SCD detector (1000 ppm). why the result it 279 ppm Best regards
While performing TOC sst analysis Zero shift disabled & sample analysis zero shift enabled why?
what parameters we will consider while developing a HPLC method and how we confirm our HPLC method is valid?
why multimedia dissolution require to do?
on saturation solubility study data how we can find out the bcs class of drug?
How to calculation of residual solvent methanol (3000 ppm) standard solution, and how many quantity used of test sample
Related substance method equivalency on control sample or spiked sample?
why xterra column require to use at higher ph?
What is importance of pH in the Heavy metal test and How it effect on the test results?
What is the calibration of uv process and preparation
for which product require to do content uniformity? what is limit of cu?
Why sometime potassium ortho phosphate mix with acetone use for sample preparation for hplc