how can you change unit of totalizer in siemens mag 5000
flow meter?
Answer Posted / juan carlos valenzuela
If the unit is not a custody transfer, resetting the
totalizers is done by the following steps:
1. Hold the TOP UP KEY (First key on the top left) for 2
seconds.
2. Press the CHANGE KEY (Fourth key on the bottom left)
twice to input the default password (1000).
3. Press the LOCK/UNLOCK KEY (Sixth key on the bottom right)
4. Now you are in setup mode!
5. Press the FORWARD KEY (Second key on the top or top
middle key) five times or until you see the RESET MODE
menu.
6. Press the LOCK/UNLOCK KEY (Sixth key on the bottom
right) to go into the menu.
7. First item, is totalizer 1 reset, Press the LOCK/UNLOCK
KEY (Sixth key on the bottom right) to reset totalizer 1.
Now by pressing the FORWARD KEY (Second key on the top or
top middle key) once, you now have access to totalizer 2
reset. Follow the same procedure as in step 7.
8. After resetting the totalizer Hold the TOP UP KEY (First
key on the top left) for 2 seconds, to go back to the
operator screen.
The shortcut.......If available on the operator screen...
Press the FORWARD KEY (Second key on the top or top middle
key) until you see TOTALIZER RESET, Press the LOCK/UNLOCK
KEY (Sixth key on the bottom right), Totalizer 1 or 2 are
now reset...
Please refer to below link for the complete instruction
manual
http://support.automation.siemens.com/WW/livelink.exe?
func=cslib.csinfo&lang=en&siteid=csius&aktprim=0&extranet=st
andard&viewreg=WW&objid=16878972&treeLang=en
If you have any questions please contact the Siemens
Technical Support Line @ 1800 333 7421
| Is This Answer Correct ? | 1 Yes | 2 No |
Post New Answer View All Answers
how to check earthing,what is minum ohms for proper earthing
Draw a P& ID showing – orifice flange, plate, dp transmitter, isolation & bypass valves and tubing.
what is the input of digital counter during the the calibration time?
, Instrument Index, P&ID review, I/O schedule, datasheet, DCS / Analyzers specifications, loop diagrams, hook-up drawings.
sir i m preparing for BARC but i am not getting any papers to prepare for this exam so please send me some sample papers to my i.d for the preparation of barc.
Explain why use mmwc unit?
how to measure oxygen which methods are using?
why we take %i0.0 as input for twido plc
what is function of switch, and how is it work, explain with difination
what is diffrance between FF & SMART, explain with difination
what is the imbustion control loop
how to calibrate steam control valve in steam turbine
What is exact difference between the panel earthing and system earthing in DCS system. What is the value of resistence.
dear sir we have insatlled positioner model no.sipart PS2 make Siemens at our control valves . now problem is if we want to increase the response of valve opening speedy /fast. what parameters do we change for positioner ? any idea. please guide us.
I will attend an interview with GasGas company for a position of Instrument Technician ( Offshore )in next month can anyone tell me what questions they will ask. Thanks in advance.