How to calculate LOD & LOQ in HPLC Validation Method
Answer Posted / abdulkalim
Based on standard deviation of the response and the slope
obtained will be calculated as:
Limit of Detection (LOD):
DL= 3(SD/Slope)
Limit of quantification(LOQ):
DL=10(SD/Slope)
| Is This Answer Correct ? | 231 Yes | 20 No |
Post New Answer View All Answers
WHAT IS THE CRITERIA FOR SELECTION OF TIME OF DISSOLUTION AND THE MEDIUM OF DISSOLUTION?
what is the difference between potentiometric titration and karl fischer titration?
Please give idea about method development for cleaning method and how maco level establish
before starting analytical method valodation what you checking? and how giving preference to start validation?
USP methodology, EP methodology, IP methodology, among three if possible to use one methodology for qualify working standard to use USP, EP, IP ? Please explain...
In which situation we require to prepare the standard solution from sample in Related substance method?
[3/30, 13:29] Manoj P Venkatpurwar: How hplc column selection according to structure? How mobile phase buffer selection on molecule structure?
how to select short coloum and long coloumn for new molecule
While performing TOC sst analysis Zero shift disabled & sample analysis zero shift enabled why?
My question about gas chromatography sulfur chemiluminsecence detector. I test unknown sample gas by GC-SCD (calibrated ) and the result of *H2S is 279 PPM , *but when I test the same sample with the GC-TCD (calibrated ) the value of *H2S is 2500 PPM . I'd like to inform you that both GCs are calibrated and have very good operation conditions with stable parameters . the question is if the sample gas with higher H2S over detection limits of SCD detector (1000 ppm). why the result it 279 ppm Best regards
What is the difference between spectro meter and spectro photo meter?
What is intact assay method development.....?
In which situation we require to analytical method validation of excipient?
if you given one product then which tests you will perform?
Related substance method equivalency on control sample or spiked sample?