why potassium dicromate use in calibration of UV VIIBLE
SPECTROCOPY ?
Answer Posted / vikram
k2cr2o7 in H2So4 can give more maximum in the UV-VIS region,(at 235,257,313,350,430nm)so that we can check the whole UV-VIS region with a single run.
k2cr2o7 itself is stable and available in high purity.In dilute perchloric acid solution,it has a linear response with temperature.
| Is This Answer Correct ? | 2 Yes | 0 No |
Post New Answer View All Answers
What is aggregate and fragments in SEC?
4. Describe the operation of the Craig apparatus. Chapter.3 Equilibrium Processes in Separations
how many types of balance calibration test?
which are the diffrent batches in the pharmaceuticals?
My question about gas chromatography sulfur chemiluminsecence detector. I test unknown sample gas by GC-SCD (calibrated ) and the result of *H2S is 279 PPM , *but when I test the same sample with the GC-TCD (calibrated ) the value of *H2S is 2500 PPM . I'd like to inform you that both GCs are calibrated and have very good operation conditions with stable parameters . the question is if the sample gas with higher H2S over detection limits of SCD detector (1000 ppm). why the result it 279 ppm Best regards
what is classification of elemental impurities? what is risk assement in elemental impurities?
iam usig ph buffers merk. manually how to prepare ?
HI,I CLEARED BOB CLERK EXAM. MY INTERVIEW WILL BE ON 9TH OCTOBER,2010.PLEASE SEND ME INTERVIEW QUESTIONS AND ANSWERS. THANK YOU.
mode of absorption in alimentary canal?
what is definition of validation? which components are followed give detail?
what parameters we will consider while developing a HPLC method and how we confirm our HPLC method is valid?
can i use hplc detector to uplc and why?
What is mean by PDR?
WHY given much noise PDA dector then VWD Dector
Home heating oil or furnace oil what is made from? Can (base oil plus diesel/kerosen) = home heating oil?