we calibrate HPLC once in 6 months while iam doing
calibration i came to know that calibration was failed.
then how can i justify that the samples analyzed on the
system during this 6 months was wright? how can i forward in
to the next step for those batches?

Answer Posted / k l reddy

Need not to analyze all previous batches but you should
analyze 6 previous batches(recent batches of before
calibration)on different system(which is qualified) and
compare the results with old results. Fallow the same
prociser until the resukts matched with old results.

Is This Answer Correct ?    2 Yes 2 No



Post New Answer       View All Answers


Please Help Members By Posting Answers For Below Questions

WHAT IS THE PURPOSE WE USE HCL, CHOPPER, BURNER, MONOCHROMATOR in AAS? PLEASE EXPLAIN BRIEFLY

2121


Did anybody have method for acetyl cysteine effervescnce tablet

1831


How to choose the which salt is suitable for mobilephase

2669


in which situation require to use paddle and basket?

301


Tell me something about Vitamin A test method by HPLC

2218






How to determine water content of bis tetrazole amine mono ammonium salt

2982


What is aggregate and fragments in SEC?

1597


What is diffrence between extractable volume and deliverable volume? Answer pls

3963


In HPLC calibration, caffeine is used as primary standard for wave length calibration due to caffeine is having dual maxima at 273 & 205 nm and one minima at 245 nm. Any body can give reference of these details from any pharmacopeia (with chapter no.) or any other authentic guideline?

1469


How we performed the force degradation for drug substance, is any specific guideline is available for each parameter(Acidic, basic, oxidation,heat)? what conditions you mentained for above parameters.

1804


How to start the dissolution development for unknown tab?

1158


How to set analytical specification for combination products?

334


How will u perform degradation study in detectors other thann PDA in HPLC Methodology

2030


WHY pKa value is more important than pH value of the mobile phase in HPLC?

1267


UV and PDA detector, which have less signal to noise ratio?

657